The Effect of Aerosol Devices and Administration Techniques on Drug Delivery in a Simulated Spontaneously Breathing Pediatric Model with a Tracheostomy
نویسندگان
چکیده
Background: Evidence on aerosol delivery via tracheostomy is lacking. The purpose of this study was to evaluate the effect of aerosol device and administration technique on drug delivery in a simulated spontaneously breathing pediatric model with tracheostomy. Methods: Delivery efficiencies during spontaneous breathing with assisted and unassisted administration techniques were compared using the jet nebulizer (JNMicroMist), vibrating mesh nebulizer (VMNAeroneb Solo) and pressurized metered-dose inhaler (pMDIProAirHFA). The direct administration of aerosols in spontaneously breathing patients (unassisted technique) was compared to administration of aerosol therapy via a manual resuscitation bag (assisted technique) attached to the aerosol delivery device and synchronized with inspiration. An in-vitro lung model consisted of an uncuffed tracheostomy tube (4.5 mmID) was attached to a collecting filter (Respirgard) which was connected to a dual-chamber test lung (TTL) and a ventilator (Hamilton). The breathing parameters of a 2 years-old child were set at an RR of 25 breaths/min, a Vt of 150 mL, a Ti of 0.8 sec and PIF of 20 L/min. Albuterol sulfate was administered with each nebulizer (2.5 mg/3 ml) and pMDI with spacer (4 puffs, 108 μg/puff). Each aerosol device was tested five times with both administration techniques (n=5). Drug collected on the filter was eluted with 0.1 N HCl and analyzed via spectrophotometry. Results: The amount of aerosol deposited in the filter was quantified and expressed as inhaled mass and inhaled mass percent. The pMDI with spacer had the highest inhaled mass percent, while the VMN had the highest inhaled mass. The results of this study also found that JN had the least efficient aerosol device used in this study. The trend of higher deposition with unassisted versus assisted administration of aerosol was not significant (p>0.05). Conclusions: Drug deposited distal to the tracheostomy tube with JN was lesser than either VMN or pMDI. Delivery efficiency was similar with unassisted and assisted aerosol administration technique in this in vitro pediatric model. THE EFFECT OF AEROSOL DEVICES AND ADMINISTRATION TECHNIQUES ON DRUG DELIVERY IN A SIMULATED SPONTANEOUSLY BREATHING PEDIATRIC MODEL WITH A TRACHEOSTOMY By Bshayer R. Alhamad A Thesis Presented in Partial Fulfillment of Requirements for the Degree of Masters of Science in Health Sciences in the Division of Respiratory Therapy in Byrdine F. Lewis School of Nursing and Health Professions Georgia State University Atlanta, Georgia 2013
منابع مشابه
Effect of Aerosol Devices and Administration Techniques on Drug Delivery in a Simulated Spontaneously Breathing Pediatric Tracheostomy Model.
BACKGROUND This study was conducted to compare the efficiency of jet nebulizers, vibrating mesh nebulizers, and pressurized metered-dose inhalers (pMDI) during assisted and unassisted administration techniques using a simulated spontaneously breathing pediatric model with a tracheostomy tube (TT). METHODS An in vitro breathing model consisting of an uncuffed TT (4.5-mm inner diameter) was att...
متن کاملNebulized albuterol delivery in a model of spontaneously breathing children with tracheostomy.
BACKGROUND Nebulized therapy is commonly used in spontaneously breathing tracheostomized patients, despite a lack of recommended devices and techniques. I compared albuterol dose delivered to a model of spontaneously breathing children with tracheostomy, using different nebulizers, tracheostomy tube sizes, inhalation techniques, and breathing patterns. METHODS A tracheostomy model was connect...
متن کاملQuantifying Aerosol Delivery in Simulated Spontaneously Breathing Patients With Tracheostomy Using Different Humidification Systems With or Without Exhaled Humidity.
BACKGROUND Aerosol and humidification therapy are used in long-term airway management of critically ill patients with a tracheostomy. The purpose of this study was to determine delivery efficiency of jet and mesh nebulizers combined with different humidification systems in a model of a spontaneously breathing tracheotomized adult with or without exhaled heated humidity. METHODS An in vitro mo...
متن کاملComparison of HFNC, bubble CPAP and SiPAP on aerosol delivery in neonates: An in-vitro study.
UNLABELLED Aerosol drug delivery via high flow nasal cannula (HFNC), bubble continuous positive airway pressure (CPAP), and synchronized inspiratory positive airway pressure (SiPAP) has not been quantified in spontaneously breathing premature infants. OBJECTIVES The purpose of this study was to compare aerosol delivery via HFNC, bubble CPAP, and SiPAP in a model of a simulated spontaneously b...
متن کاملPerformance Comparisons of Jet and Mesh Nebulizers Using Different Interfaces in Simulated Spontaneously Breathing Adults and Children.
BACKGROUND Different types of nebulizers and interfaces are used for the treatment of adults and children with pulmonary diseases. The purpose of this study was to determine the efficiency of a mesh nebulizer (MN) with a proprietary adapter and a jet nebulizer (JN) under different configurations in adult and pediatric models of spontaneous breathing. We hypothesize that delivery efficiency of J...