Performance of iterative image reconstruction in CT of the paranasal sinuses: a phantom study.
نویسندگان
چکیده
BACKGROUND AND PURPOSE CT in low dose technique is the criterion standard imaging modality for evaluation of the paranasal sinus. Our aim was to evaluate the dose-reduction potential of a recently available sinogram-affirmed iterative reconstruction technique, regarding noise, image quality, and time duration when evaluating this region. MATERIALS AND METHODS CT was performed on a phantom head at different tube voltages (120 kV, 100 kV) and currents (100 mAs, 50 mAs, 25 mAs). Each protocol was reconstructed (in soft tissue and bony kernel) by using standard filtered back-projection and 5 different SAFIRE strengths, and image noise was evaluated. Subjective image quality was evaluated on noise-aligned image triplets acquired at tube currents of 100% (FBP), 50% (SAFIRE), and 25% (SAFIRE) by using a 5-point scale (1 = worst, 5 = best). The time duration for image reconstruction was noted for calculations with FBP and SAFIRE. RESULTS SAFIRE reduced image noise by 15%-85%, depending on the iterative strength, rendering kernel, and dose parameters. Noise reduction was stronger at a bone kernel algorithm both in 1- and 3-mm images (P < .05). Subjective quality evaluation of the noise-adapted images showed preference for those acquired at 100% tube current with FBP (4.7-5.0) versus 50% dose with SAFIRE (3.4-4.4) versus 25% dose with SAFIRE (2.0-3.1). The time duration for FBP image sets was 2.9-6.6 images per second versus SAFIRE with 0.9-1.6 images per second. CONCLUSIONS For CT of the paranasal sinus, SAFIRE algorithms are suitable for image-noise reduction. Because image quality decreases with dosage, careful choice of the appropriate iterative method is necessary to achieve an optimal balance between image noise and quality.
منابع مشابه
Image Optimization and reduction of radiation dose in CT of the paranasal sinuses
Introduction: Since demand for computed tomography (CT) examinations is growing, patients are exposed more frequently to ionizing radiation. To reduce the dose delivered to patients in each scan, CT technologists can change the image acquisition parameters. However, this reduces image quality. Present study aimed to optimize and reduce radiation dose in a CT of the paranasal si...
متن کاملComparing IDREAM as an Iterative Reconstruction Algorithm against In Filtered Back Projection in Computed Tomography
Introduction: Recent studies of Computed Tomography (CT) conducted on patient dose reduction have recommended using an iterative reconstruction algorithm and mA (mili-Ampere) dose modulation. The current study aimed to evaluate Iterative Dose Reduction Algorithm (IDREAM) as an iterative reconstruction algorithm. Material and Methods: Two CT p...
متن کاملFast System Matrix Calculation in CT Iterative Reconstruction
Introduction: Iterative reconstruction techniques provide better image quality and have the potential for reconstructions with lower imaging dose than classical methods in computed tomography (CT). However, the computational speed is major concern for these iterative techniques. The system matrix calculation during the forward- and back projection is one of the most time- cons...
متن کاملTailored low dose three-dimensional CT of paranasal sinuses.
A technique for three-dimensional (3D) computed tomography (CT) of the paranasal sinuses was evaluated. The influence of milliamperage and reconstruction algorithms on image quality was studied in a phantom. Eleven patients with bony abnormalities of the paranasal sinuses underwent CT to assess the clinical impact of 3D CT protocol. Contiguous 1.5-mm sections obtained at 120 kVp/175 mA/1.9 sec ...
متن کاملAnatomic Reference for Computed Tomography of the Paranasal Sinuses and Their Openings in the Rayini Goat
Objective- to provide a reference anatomy of the paranasal sinuses of the goat by using computed tomographic modality. Design- Experimented study. Animals- Five male Rayini goats. Procedures- CT images were acquired from the head region perpendicular to the hard palate. CT windows were adjusted as necessary to obtain the optimal image for the paranasal sinuses. The images were studie...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- AJNR. American journal of neuroradiology
دوره 34 5 شماره
صفحات -
تاریخ انتشار 2013