Detecting Triple-Vessel Disease with Cadmium Zinc Telluride-Based Single-Photon Emission Computed Tomography Using the Intensity Signal-to-Noise Ratio between Rest and Stress Studies
نویسندگان
چکیده
The purpose of this study was to investigate if a novel parameter, the stress-to-rest ratio of the signal-to-noise ratio (RSNR) obtained with a cadmium zinc telluride (CZT) SPECT scanner, could be used to distinguish triple-vessel disease (TVD) patients. Methods. One hundred and two patients with suspected coronary artery disease were retrospectively involved. Each subject underwent a Tl-201 SPECT scan and subsequent coronary angiography. Subjects were separated into TVD (n = 41) and control (n = 61) groups based on coronary angiography results using 50% as the stenosis cutoff. The RSNR was calculated by dividing the stress signal-to-noise ratio (SNR) by the rest SNR. Summed scores were calculated using quantitative perfusion SPECT (QPS) for all subjects. Results. The RSNR in the TVD group was found to be significantly lower than that in the control group (0.83 ± 0.15 and 1.06 ± 0.17, resp.; P < 0.01). Receiver-operating characteristic (ROC) analysis showed that RSNR can detect TVD more accurately than the summed difference score with higher sensitivity (85% versus 68%), higher specificity (90% versus 72%), and higher accuracy (88% versus 71%). Conclusion. The RSNR may serve as a useful index to assist the diagnosis of TVD when a fully automatic quantification method is used in CZT-based SPECT studies.
منابع مشابه
Comparison between ultrafast and standard single-photon emission CT in patients with coronary artery disease: a pilot study.
BACKGROUND A novel technology has been developed for ultrafast (UF) single-photon emission CT (SPECT) myocardial perfusion imaging by using a pinhole collimation design and multiple cadmium zinc telluride crystal arrays. The purpose of this study was to compare myocardial perfusion imaging obtained by UF-SPECT with standard (S) SPECT in patients with known or suspected coronary artery disease. ...
متن کاملSingle-scan rest/stress imaging with 99mTc-Sestamibi and cadmium zinc telluride-based SPECT for hyperemic flow quantification: A feasibility study evaluated with cardiac magnetic resonance imaging
INTRODUCTION We aimed to evaluate whether the hyperemic myocardial blood flow (MBF) can be estimated using cadmium zinc telluride (CZT)-based single-photon emission computed tomography (SPECT) cameras with a single, rapid rest/stress dynamic scan. Dynamic contrast-enhanced (DCE) cardiac magnetic resonance imaging (MRI) was used as a reference modality for flow measurement. MATERIALS AND METHO...
متن کاملComparison Between Ultrafast and Standard SPECT in Patients with Coronary Artery Disease: A Pilot Study
Background A novel technology has been developed for ultrafast (UF) singlephoton emission computed tomography (SPECT) myocardial perfusion imaging by employing a pin-hole collimation design and multiple cadmium zinc telluride crystal arrays. The purpose of this study was to compare myocardial perfusion imaging obtained by UF-SPECT with standard (S) SPECT in patients with known or suspected coro...
متن کاملEvaluation of a Wobbling Method Applied to Correcting Defective Pixels of CZT Detectors in SPECT Imaging
In this paper, we propose a wobbling method to correct bad pixels in cadmium zinc telluride (CZT) detectors, using information of related images. We build up an automated device that realizes the wobbling correction for small animal Single Photon Emission Computed Tomography (SPECT) imaging. The wobbling correction method is applied to various constellations of defective pixels. The corrected i...
متن کاملCombined supine and prone myocardial perfusion single-photon emission computed tomography with a cadmium zinc telluride camera for detection of coronary artery disease.
BACKGROUND Myocardial perfusion SPECT (MPS) traditionally requires the patient to be in the supine position, but diaphragmatic attenuation of the inferior wall reduces test specificity. The aim of this study was to assess the feasibility of combined MPS in the supine and prone positions using a novel cadmium zinc telluride (CZT) camera. METHODS AND RESULTS A total of 276 consecutive patients ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
دوره 2017 شماره
صفحات -
تاریخ انتشار 2017