Breakage of the Tail Portion of the Lag Screw during Removal of Proximal Femoral Zimmer Natural Nail: Report of Two Cases with Technical Notes

نویسندگان

  • Asep Santoso
  • Ik-Sun Choi
  • Kyung-Soon Park
  • Taek-Rim Yoon
چکیده

Proximal femoral cephalomedullary nailing is one of the treatments of choice for intertrochanteric fracture. Zimmer Natural Nail® (ZNN; Zimmer) is one of the alternatives. We report two cases of broken tail portion of the ZNN lag screw during the removal procedure. This may be the first reported cases in scientific literature with this pattern of failure. We report these cases to highlight one of the possible problems that could occur during removal of this implant which can possibly prolong the operation time significantly if the required device is not prepared.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Gamma nail breakage: a report of four cases.

Gamma nails have been used extensively for the treatment of proximal femoral fractures. Nail breakage at the level of the aperture of the lag screw is rare. We report 4 such cases mainly associated with a large posteromedial cortex gap and nonunion. The need for adequate reduction to avoid such a complication is emphasised.

متن کامل

Type II Intertrochanteric Fractures: Proximal Femoral Nailing (PFN) Versus Dynamic Hip Screw(DHS)

  Background: Intertrochanteric fracture is one of the most common fractures of the hip especially in the elderly with osteoporotic bones, usually due to low-energy trauma like simple falls. Dynamic Hip Screw (DHS) is still considered the gold standard for treating intertrochanteric fractures by many. Not many studies compare the DHS with Proximal femoral nail (PFN), in Type II intertrochanteri...

متن کامل

Breakage of a Lag Screw of Cephalomedullary Nail: A Technique of Removal

A broken lag screw of the cephalomedullary nail is a rare condition. Removal of the retained lag screw from the femoral head is also very challenging. This article describes a surgical technique and the modified instrument that was available in the operating room for removing the broken implant by closed technique.

متن کامل

Numerical Optimization of the Position in Femoral Head of Proximal Locking Screws of Proximal Femoral Nail System; Biomechanical Study

BACKGROUND Proximal femoral fracture rates are increasing due to osteoporosis and traffic accidents. Proximal femoral nails are routinely used in the treatment of these fractures in the proximal femur. AIMS To compare various combinations and to determine the ideal proximal lag screw position in pertrochanteric fractures (Arbeitsgemeinschaft für Osteosynthesefragen classification 31-A1) of th...

متن کامل

Intrapelvic Protrusion of a Broken Guide Wire Fragment during Fixation of a Femoral Neck Fracture

Breakage of DHS guide wire during surgery and its migration into the pelvis through the hip joint is a rare complication and its removal can be very challenging for the surgeon. We share our experience of a similar case wherein we used an ‘iliofemoral’ approach to successfully remove the broken transfixing guidewire from the hip joint. Although iliofemoral approach is similar to the lateral win...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 29  شماره 

صفحات  -

تاریخ انتشار 2017