High incidence of periprosthetic joint infection with propionibacterium acnes after the use of a stemless shoulder prosthesis with metaphyseal screw fixation - a retrospective cohort study of 241 patients propionibacter infections after eclipse TSA

نویسندگان

  • Lisa Johansson
  • Nils P Hailer
  • Hans Rahme
چکیده

BACKGROUND A stemless shoulder prosthesis with humeral metaphyseal screw fixation was introduced in order to save bone-stock and to facilitate reconstruction of biomechanics (Eclipse®). The aim of this study was to analyze whether the risk of infection is different with this implant compared to conventional shoulder prosthesis. METHODS Two hundred and forty-one patients (54.8% females) were operated with a shoulder arthroplasty and followed for median 2.0 (0.1-5.7) years. One hundred and two (42.3%) had received an Eclipse® prosthesis, the remaining patients were operated with other implants. There was an overrepresentation of males in the Eclipse® group (63.7% males) when compared with the control group (31.7% males). RESULTS In the Eclipse® group 10 (9.8%) patients developed a periprosthetic joint infection, as opposed to 1 (0.7%) in the control group. The most common bacteria was Propionibacterium acnes. Unadjusted infection-free survival after 4 years was 88.8% (CI 82.5-95.7) for Eclipse® patients and 95.7% (CI 87.7-100.0) for controls (p = 0.002). After adjustment for age, gender, diagnosis, and type of shoulder prosthesis (total or hemi), the risk ratio for revision due to infection was 4.3 (CI 0.5-39.1) for patients with the Eclipse® prosthesis. CONCLUSIONS Deep infections seem to be more common after the use of the metaphyseally fixed Eclipse® prosthesis than after conventional shoulder implants, but a predominance of male gender and younger age in the Eclipse group may have biased our findings. Future studies on larger cohorts and in vitro investigations on bacterial adherence and biofilm formation are needed. TRIAL REGISTRATION The study was conducted in accordance with the Helsinki Declaration. The local ethics board (Regionala Etikprövningsnämnden Stockholm) approved the study (Approval number 2015/1590-31, date of issue 2015-10-14). Retrospectively registered.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Erratum to: High incidence of periprosthetic joint infection with propionibacterium acnes after the use of a stemless shoulder prosthesis with metaphyseal screw fixation - a retrospective cohort study of 241 patients propionibacter infections after eclipse TSA

Erratum After the publication of the article [1] is was brought to our attention that the title was incorrectly given as “High incidence of periprosthetic joint infection with propionibacterium acnes after the use of a stemless shoulder prosthesis with metaphyseal screw fixation a retrospective cohort study of 241 patients propionibacter infections after eclipse TSA”. The correct title is “High...

متن کامل

Comparison of Acute Acromioclavicular Joint Dislocation Treatment by Screw Fixation with and without Ligament Reconstruction

Background & Aims: Acute acromioclavicular (AC) joint dislocation is common in shoulder injuries. Considering the biomechanics of this joint, the use of a treatment method, which can restore joint physiology and biomechanics to the extent possible, is important. This study aimed to compare the treatment of acute AC joint dislocation by screw fixation with and without ligament reconstruction. Me...

متن کامل

Radiological changes do not influence clinical mid-term outcome in stemless humeral head replacements with hollow screw fixation: a prospective radiological and clinical evaluation

BACKGROUND Stemless shoulder arthroplasty is a fairly new concept. Clinical and radiological follow-up is essential to prove implant safety and concept. This prospective single-centre study was performed to evaluate the influence of radiological changes on clinical mid-term outcome following stemless humeral head replacement with hollow screw fixation. METHODS Short- and mid-term radiological...

متن کامل

Periprosthetic Joint Infection of Shoulder Arthroplasties: Diagnostic and Treatment Options

Periprosthetic joint infection (PJI) is one of the most frequent reasons for painful shoulder arthroplasties and revision surgery of shoulder arthroplasties. Cutibacterium acnes (Propionibacterium acnes) is one of the microorganisms that most often causes the infection. However, this slow growing microorganism is difficult to detect. This paper presents an overview of different diagnostic test ...

متن کامل

First Metatarsophalangeal Joint Arthrodesis: A Retrospective Comparison of Crossed-screws, Locking and Non-Locking Plate Fixation with Lag Screw

Background:Locking plate fixation is increasingly used for first metatarsophalangeal joint (MTP-I) arthrodesis. Still there is few comparable clinical data regarding this procedure. Methods:We retrospectively evaluated 60 patients who received an arthrodesis of the MTP-I between January 2008 and June 2010. With 20 patients each we performed a locking plate fixation with lag screw, arthrodesis w...

متن کامل

ذخیره در منابع من