Perturbation of TG‐43 parameters of the brachytherapy sources under insufficient scattering materials
نویسندگان
چکیده
In the recommendations of Task Group #43 from American Association of Physicists in Medicine (AAPM TG43), methods of brachytherapy source dosimetry are recommended, under full scattering conditions. However, in actual brachytherapy procedures, sources may not be surrounded by full scattering tissue in all directions. Clinical examples include high-dose-rate (HDR) brachytherapy of the breast or low-dose-rate (LDR) brachytherapy of ocular melanoma using eye plaque treatment with 125I and 103Pd. In this work, the impact of the missing tissue on the TG-43-recommended dosimetric parameters of different brachytherapy sources was investigated. The impact of missing tissue on the TG-43-recommended dosimetric parameters of 137Cs, 192Ir, and 103Pd brachytherapy sources was investigated using the MCNP5 Monte Carlo code. These evaluations were performed by placing the sources at different locations inside a 30 × 30 × 30 cm3 cubical water phantom and comparing the results with the values of the source located at the center of the phantom, which is in a full scattering condition. The differences between the thickness of the overlying tissues for different source positions and the thickness of the overlying tissue in full scattering condition is referred to as missing tissue. The results of these investigations indicate that values of the radial dose function and 2D anisotropy function vary as a function of the thickness of missing tissue, only in the direction of the missing tissue. These changes for radial dose function were up to 5%, 11%, and 8% for 137Cs, 192Ir, and 103Pd, respectively. No significant changes are observed for the values of the dose rate constants. In this project, we have demonstrated that the TG-43 dosimetric parameters may only change in the directions of the missing tissue. These results are more practical than the published data by different investigators in which a symmetric effect of the missing tissue on the dosimetric parameters of brachytherapy source are being considered, regardless of the implant geometry in real clinical cases.
منابع مشابه
Simulation of the shielding effects of an applicator on the AAPM TG-43 parameters of CS-137 Selectron LDR brachytherapy sources
Background: The dose rate distribution delivered by a low dose rate 137Cs pellet source, a spherical source used within the source trains of the Selectron gynecological brachytherapy system, was investigated using the MCNP4C Monte Carlo code. Materials and Methods: The calculations were performed in both water and Plexiglas and the absolute dose rate distribution for a single pellet s...
متن کاملDetermination of TG-43 Dosimetric Parameters for Photon Emitting Brachytherapy Sources
Objective: Brachytherapy sources are widely used for the treatment of cancer. The report of Task Group No. 43 (TG-43) of American Association of Physicists in Medicine is known as the most common method for the determination of dosimetric parameters for brachytherapy sources. The aim of this study is to obtain TG-43 dosimetric parameters for 60Co, 137Cs, 192Ir and 103Pd brachyt...
متن کاملMaterial-specific Conversion Factors for Different Solid Phantoms Used in the Dosimetry of Different Brachytherapy Sources
Introduction Based on Task Group No. 43 (TG-43U1) recommendations, water phantom is proposed as a reference phantom for the dosimetry of brachytherapy sources. The experimental determination of TG-43 parameters is usually performed in water-equivalent solid phantoms. The purpose of this study was to determine the conversion factors for equalizing solid phantoms to water. Materials and Methods T...
متن کاملDetermination of Dosimetric characteristics of a New 192Ir-PDR Brachytherapy Source According to AAPM TG- 43 Protocol using Monte Carlo simulation technique
Introduction: 192Ir is one of the important sources frequently used in brachytherapy. Up to now, a lot of commercial models of this source have been made which Ir-192 has been recently added to them. The aim of the present study is to determine the dosimetric parameters of this new source model based on the recommendations of TG-43(U1) protocol using Monte Carlo simulation tech...
متن کاملDosimetric Parameters Estimation of I-125 Brachytherapy Source in fat phantom using GATE8.0 code
Introduction: Brachytherapy is one type of internal radiation therapy where radiation sources, which are usually encapsulated, are placed as close as possible to the tumor site or inside the patient's body. In this technique, it is important to determine dose distribution around the brachytherapy capsule to create optimal treatment plant. In this way, dosimetric parameters are...
متن کامل