Highly Stable PEDOT:PSS Coating on Gold Microelectrodes with Improved Charge Injection Capacity for Chronic Neural Stimulation

نویسنده

  • Walter Lang
چکیده

This study introduces two new processes that highly enable PEDOT:PSS (poly(3,4-ethylenedioxythiophene)-poly(styrenesulfonate)) as stable coating material for chronic neural stimulation. In first process, strong mechanical bonding between PEDOT:PSS coating and gold electrodes is achieved by creating rough porous surface with partial iodine etching. PEDOT:PSS coating on iodine etched gold electrode shows 100% stability under strong ultrasonic treatment. The second process represents electrochemical modification of PEDOT:PSS coating by cyclic voltammetry method in Ringer’s solution. This process reduces electrode polarization 33% during stimulation. Therefore, charge injection capacity increases that ensures safe stimulation. A combination of both processes facilitates the use of PEDOT:PSS coating for successful chronic neural recording and stimulation.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Demonstration of Intracortical Chronic Recording and Acute Microstimulation Using Novel Floating Neural Probes

This paper presents long-term stable multichannel recording of neural activity using novel intracortical floating probes implanted chronically in rat cortex. The novel flexible probe design approach allows recording of action potentials for at least 38 days after implantation. Furthermore the capability of the PEDOT: PSS coated microelectrodes for electrical stimulation is characterized in vitr...

متن کامل

Scaling Effects on the Electrochemical Stimulation Performance of Au, Pt, and PEDOT:PSS Electrocorticography Arrays

The efficacy of electrical brain stimulation in combatting neurodegenerative diseases and initiating function is expected to be significantly enhanced with the development of smaller scale microstimulation electrodes and refined stimulation protocols. These benefits cannot be realized without a thorough understanding of scaling effects on electrochemical charge injection charac­ teristics. This...

متن کامل

Organic electrode coatings for next-generation neural interfaces

Traditional neuronal interfaces utilize metallic electrodes which in recent years have reached a plateau in terms of the ability to provide safe stimulation at high resolution or rather with high densities of microelectrodes with improved spatial selectivity. To achieve higher resolution it has become clear that reducing the size of electrodes is required to enable higher electrode counts from ...

متن کامل

Multifunctional Nanobiomaterials for Neural Interfaces

Neural electrodes are designed to interface with the nervous system and provide control signals for neural prostheses. However, robust and reliable chronic recording and stimulation remains a challenge for neural electrodes. Here, a novel method for the fabrication of soft, low impedance, high charge density, and controlled releasing nanobiomaterials that can be used for the surface modificatio...

متن کامل

Gold nanograin microelectrodes for neuroelectronic interfaces.

Neuroelectronic interfaces are imperative in investigating neural tissues as electrical signals are the main information carriers in the nervous system and metal microelectrodes have been widely used for recording and stimulation of nerve cells. For high performance microelectrodes, low tissue-electrode interfacial impedance and high charge injection limits are essential and nanoscale surface e...