Torsional ultrasound mode versus combined torsional and conventional ultrasound mode phacoemulsification for eyes with hard cataract

نویسندگان

  • Mohamed A Fakhry
  • Malak I El Shazly
چکیده

PURPOSE To compare torsional versus combined torsional and conventional ultrasound modes in hard cataract surgery regarding ultrasound energy and time and effect on corneal endothelium. SETTINGS Kasr El Aini hospital, Cairo University, and International Eye Hospital, Cairo, Egypt. METHODOLOGY Ninety-eight eyes of 63 patients were enrolled in this prospective comparative randomized masked clinical study. All eyes had nuclear cataracts of grades III and IV using the Lens Opacities Classification System III (LOCS III). Two groups were included, each having an equal number of eyes (49). The treatment for group A was combined torsional and conventional US mode phacoemulsification, and for group B torsional US mode phacoemulsification only. Pre- and post-operative assessments included best corrected visual acuity (BCVA), intraocular pressure (IOP), slit-lamp evaluation, and fundoscopic evaluation. Endothelial cell density (ECD) and central corneal thickness (CCT) were measured preoperatively, 1 day, 7 days, and 1 month postoperatively. All eyes were operated on using the Alcon Infiniti System (Alcon, Fort Worth, TX) with the quick chop technique. All eyes were implanted with AcrySof SA60AT (Alcon) intraocular lens (IOL). The main phaco outcome parameters included the mean ultrasound time (UST), the mean cumulative dissipated energy (CDE), and the percent of average torsional amplitude in position 3 (%TUSiP3). RESULTS Improvement in BCVA was statistically significant in both groups (P < 0.001). Comparing UST and CDE for both groups revealed results favoring the pure torsional group (P = 0.002 and P < 0.001 for UST; P = 0.058 and P = 0.009 for CDE). As for %TUSiP3, readings were higher for the pure torsional group (P = 0.03 and P = 0.01). All changes of CCT, and ECD over time were found statistically significant using one-way ANOVA testing (P < 0.001). CONCLUSION Both modes are safe in hard cataract surgery, however the pure torsional mode showed less US energy used.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Efficacy of Torsional Phaco in Hard Egyptian Cataract Versus Combined Torsional and Longitudinal Phaco Using the Infiniti Vision System

Purpose: To compare the efficacy and phacoemulsification parameters of torsional mode versus the combined torsional and longitudinal ultrasound mode in hard cataract of Egyptian patients. Methods: This randomized comparative study evaluated eyes 1, 7, and 30 days after phacoemulsification with an Infiniti Vision System using the torsional mode versus the combined torsional (100%) and longitudin...

متن کامل

The Comparison between Torsional and Conventional Mode Phacoemulsification in Moderate and Hard Cataracts

PURPOSE To compare the intraoperative performances and postoperative outcomes of cataract surgery performed with longitudinal phacoemulsification and torsional phacoemulsification in moderate and hard cataracts. METHODS Of 85 patients who had senile cataracts, 102 eyes were operated on using the Infiniti Vision System. Preoperative examinations (slit lamp examination, mean central corneal thi...

متن کامل

Torsional ultrasound modality for hard nucleus phacoemulsification cataract extraction

AIM To evaluate the efficacy and safety of phacoemulsification using torsional modality with different parameter settings for hard nucleus cataract extraction. DESIGN A prospective, randomised clinical study. METHODS A clinical practice study conducted at the Cataract Service, Zhongshan Ophthalmic Center, Sun-Yat-Sen University, and Guangzhou. One eye each from 198 consecutive patients with...

متن کامل

Fluid dynamics, cavitation, and tip-to-tissue interaction of longitudinal and torsional ultrasound modes during phacoemulsification.

PURPOSE To describe the fluidic events that occur in a test chamber during phacoemulsification with longitudinal and torsional ultrasound (US) modalities. SETTING Pasteur Ophthalmic Clinic Phacodynamics Laboratory, Santiago, Chile, and Nanyang Technological University, Singapore. DESIGN Experimental study. METHODS Ultra-high-speed videos of a phacoemulsifying tip were recorded while the t...

متن کامل

Corneal endothelial morphology and function after torsional and longitudinal ultrasound mode phacoemulsification.

PURPOSE To study the endothelial cell morphology and corneal thickness changes after phacoemulsification by using the OZil torsional and longitudinal ultrasound techniques (Infiniti Vision System, Alcon Laboratories). SETTING Department of Ophthalmology, Clinical Center, University of Debrecen, Debrecen, Hungary. METHODS 52 patients with cataract were randomly assigned to longitudinal ultra...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 5  شماره 

صفحات  -

تاریخ انتشار 2011