Effect of prosthetic valve malfunction on the Doppler-catheter gradient relation for bileaflet aortic valve prostheses.
نویسندگان
چکیده
BACKGROUND Considerable discrepancies between Doppler and catheter gradients caused by localized gradients and pressure recovery have been reported for normal bileaflet aortic valve prostheses. METHODS AND RESULTS To examine whether this Doppler-catheter gradient relation is affected by prosthetic valve malfunction, a 19-mm CarboMedics aortic valve was simultaneously studied with continuous-wave Doppler and catheter technique in normal function and in various states of malfunction ranging from slightly restricted opening to total occlusion of one leaflet. For each functional status, peak and mean gradients were measured at eight different flow rates (cardiac output, 2.0-6.0 L/min). Excellent correlation between Doppler and catheter gradients was found regardless of the valve function (r = 0.99, SEE = 1.0-3.3 mm Hg). However, the relation between Doppler and catheter gradient was highly dependent on the function of the valve as shown by a variation of slopes from 1.08 to 2.08. For the normally functioning valve (angle between flow axis and leaflet 5 degrees), peak and mean Doppler gradients were approximately twice the catheter gradients (slope, 2.08 and 2.03 for peak and mean gradients, respectively). Slightly restricted opening of one leaflet (22 degrees) significantly altered the Doppler-catheter gradient relation, and slopes decreased to 1.69 (p < 0.01) and 1.52 (p < 0.001) for peak and mean gradients, respectively. The differences between Doppler and catheter gradients significantly decreased with further restriction of valve opening, and slopes ranged from 1.25 to 1.41 for angles between 34 degrees and 52 degrees. When one leaflet was totally occluded, the slope finally dropped to 1.08 for both peak and mean gradients, and Doppler gradients were only slightly greater than catheter gradients. Gradients increased with malfunction of the valve caused by reduction of the effective orifice area. However, the increase of Doppler gradients was considerably smaller than the increase of simultaneous catheter gradients. CONCLUSIONS The discrepancies between Doppler and catheter gradients that have been reported for normally functioning bileaflet aortic valve prostheses may be reduced or even disappear in patients with malfunctioning valves. Furthermore, the increase of Doppler gradients caused by malfunction of the valve may underestimate the true hemodynamic changes.
منابع مشابه
Doppler-Catheter Gradient Relation for Bileaflet Aortic Valve Prostheses
Background. Considerable discrepancies between Doppler and catheter gradients caused by localized gradients and pressure recovery have been reported for normal bileaflet aortic valve prostheses. Methods and Results. To examine whether this Doppler-catheter gradient relation is affected by prosthetic valve malfunction, a 19-mm CarboMedics aortic valve was simultaneously studied with continuous-w...
متن کاملContinuous wave Doppler echocardiographic measurement of prosthetic valve gradients. A simultaneous Doppler-catheter correlative study.
Studies correlating prosthetic valve gradients determined by continuous wave Doppler echocardiography with gradients obtained by cardiac catheterization have, to date, been limited to patients with mitral and tricuspid prostheses or have compared nonsimultaneous measurements. Simultaneous Doppler and catheter pressure gradients in 36 patients (mean age, 63 +/- 13 years) with 42 prosthetic valve...
متن کاملContinuous Wave Doppler Echocardiographic Measurement of Prosthetic Valve Gradients
Studies correlating prosthetic valve gradients determined by continuous wave Doppler echocardiography with gradients obtained by cardiac catheterization have, to date, been limited to patients with mitral and tricuspid prostheses or have compared nonsimultaneous measurements. Simultaneous Doppler and catheter pressure gradients in 36 patients (mean age, 63+13 years) with 42 prosthetic valves (2...
متن کاملSuccessful Use of Two Thrombolytic Drugs in Prosthetic Mitral and Aortic Valve Thrombosis
Introduction: Prosthetic valve thrombosis is a rare and severe complication of valve replacement, most often encountered with a mechanical prosthesis. The significant morbidity and mortality associated with this condition warrant rapid diagnostic evaluation. Although surgery is the first-line therapy in symptomatic obstructive mechanical valve thrombosis, thrombolytic therapy has been used as a...
متن کاملEvaluation of Björk-Shiley prosthetic valves by real-time two-dimensional Doppler echocardiographic flow mapping.
We studied the value of two-dimensional Doppler echocardiographic color flow mapping for identifying normal transvalve flow profiles and valve malfunction in 20 patients with Björk-Shiley prosthetic valves. Seven patients had Björk-Shiley prosthetic valves in the aortic position alone, seven in the mitral position, and six had prosthetic valves in both the aortic and mitral positions. In 10 pat...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Circulation
دوره 87 4 شماره
صفحات -
تاریخ انتشار 1993