Radial dependence of lineal energy distribution of 290-MeV/u carbon and 500-MeV/u iron ion beams using a wall-less tissue-equivalent proportional counter
نویسندگان
چکیده
Using a wall-less tissue-equivalent proportional counter for a 0.72-μm site in tissue, we measured the radial dependence of the lineal energy distribution, yf(y), of 290-MeV/u carbon ions and 500-MeV/u iron ion beams. The measured yf(y) distributions and the dose-mean of y, [Formula: see text], were compared with calculations performed with the track structure simulation code TRACION and the microdosimetric function of the Particle and Heavy Ion Transport code System (PHITS). The values of the measured [Formula: see text] were consistent with calculated results within an error of 2%, but differences in the shape of yf(y) were observed for iron ion irradiation. This result indicates that further improvement of the calculation model for yf(y) distribution in PHITS is needed for the analytical function that describes energy deposition by delta rays, particularly for primary ions having linear energy transfer in excess of a few hundred keV μm(-1).
منابع مشابه
Microdosimetric evaluation of secondary particles in a phantom produced by carbon 290 MeV/nucleon ions at HIMAC.
Microdosimetric single event spectra as a function of depth in a phantom for the 290 MeV/nucleon therapeutic carbon beam at HIMAC were measured by using a tissue equivalent proportional counter (TEPC). Two types of geometries were used: one is a fragment particle identification measurement (PID-mode) with time of flight (TOF) method without a backward phantom, and the other is an in-phantom mea...
متن کاملMonte Carlo computation of dose deposited by carbon ions in radiation therapy
Background: High-velocity carbon ion beams represent the most advanced tool for radiotherapy of deep-seated tumors. Currently, the superiority of carbon ion therapy is more prominent on lung cancer or hepatomas. Materials and Methods: The data for lateral straggling and projected range of monoenergetic 290 MeV/u (3.48 GeV) carbon ions in muscle tissue were obtained from the stopping and range o...
متن کاملNeutron Energy Spectra and Dose Equivalent Rates from Heavy-ion Reactions below 20 Mev/u Using the Phits Code *
Neutron energy spectra from 6.25 MeV/u C ions incident on a copper target, 10 MeV/u C ions incident carbon, copper, and lead targets, and O ions incident on a copper target have been calculated using the Quantum Molecular Dynamics model (QMD) [1] coupled to the Generalized Evaporation Model (GEM) [2] in the Particle and Heavy Ion Transport Code (PHITS) [3]. In particular, the influence of the “...
متن کاملSecondary Particles Produced by Hadron Therapy
Introduction Use of hadron therapy as an advanced radiotherapy technique is increasing. In this method, secondary particles are produced through primary beam interactions with the beam-transport system and the patient’s body. In this study, Monte Carlo simulations were employed to determine the dose of produced secondary particles, particularly neutrons during treatment. Materials and Methods I...
متن کاملThe response of a spherical tissue-equivalent proportional counter to iron particles from 200-1000 MeV/nucleon.
The radiation environment on board the space shuttle and the International Space Station includes high-Z and high-energy (HZE) particles that are part of the galactic cosmic radiation (GCR) spectrum. Iron-56 particles are considered to be one of the most biologically important parts of the GCR spectrum. Tissue-equivalent proportional counters (TEPCs) are used as active dosimeters on manned spac...
متن کامل