Late Paravalvular Aortic Regurgitation: Migration of the Valve or Late Recoil?
نویسندگان
چکیده
A 79-year-old man underwent trans-catheter aortic valve replacement for symptomatic severe aortic stenosis with a 26-mm Edwards SAPIEN XT valve. Immediately after valve deployment there was moderate amount of paravalvular leak. Post-dilation was performed with an additional 2 cc of volume, and the paravalvular leak was reduced to trace. Nine months later, trans-thoracic echocardiography revealed moderate to severe paravalvular leak and possible aortic migration of the valve. The patient was brought back for the treatment of the paravalvular leak which was suspected to be due to valve migration. However, fluoroscopy and trans-esophageal echocardiography showed good valve position. Measurement of late valve recoil in the Coplanar view using cine-angiographic analysis software showed that the lower third of the valve had the greatest late recoil (-1.74 mm, 6.55%), which presumably accounted for the progression of the paravalvular leak. Valve-in-valve trans-catheter aortic valve replacement was performed with a 26-mm SAPIEN 3 valve and the paravalvular leak was reduced to trace. This case displays late recoil as a likely mechanism for development of paravalvular leak after SAPIEN XT valve implantation. Our case illustrates that late recoil needs to be systematically evaluated in future studies, especially when trans-catheter aortic valve replacement is being expanded to lower risk and younger patients for whom the longevity and long-term performance of these valves is of critical importance.
منابع مشابه
AN INNOVATIVE SURGIC AL TECHNIQUE FOR CONTINUOUS SUTURING OF PROSTHESIS IN MITR AL VALVE REPL A CEMENT
Mitral valve replacement (MVR) is usually performed in two methods of running and interrupted suturing. In running method, the suturing of the valve is usually with some traction on the annulus and also there is a risk of aortic valve injury. This study was conducted to evaluate a new technique for suturing to avoid these complications and decrease the time of cardiac arrest. One hundred an...
متن کاملEndocarditis after transcatheter aortic valve implantation: a current assessment
Transcatheter aortic valve implantation (TAVI) constitutes an established treatment in inoperable or high perioperative risk patients with severe aortic stenosis, demonstrating similar mortality rates (at 30 days and 1 year) with surgical aortic valve replacement (SAVR). Various complications have been reported during TAVI, weeks or months post procedure. The most frequent causes of transcathet...
متن کاملLate device embolization in a persistent mitral paravalvular leak.
An 83-year-old man with severe organic mitral regurgitation underwent mitral valve surgery with implantation of a biologic prosthesis. Four months later he presented with hemolytic anemia and heart failure due to severe paravalvular regurgitation. Since the patient refused surgery, the paravalvular leak was closed percutaneously using two Amplatzer devices, with angiographic and clinical succes...
متن کاملTranscatheter Aortic Valve Replacement: A Review Article
Transcatheter aortic valve replacement (TAVR) is a novel therapeutic intervention for the replacement of severely stenotic aortic valves in high-risk patients for standard surgical procedures. Since the initial PARTNER trial results, use of TAVR has been on the rise each year. New delivery methods and different valves have been developed and modified in order to promote the minimally invasive p...
متن کاملA rare case of suture material obstructing the closure mechanism of a prosthetic aortic valve: a case report
Prosthetic aortic valve dysfunction presenting as aortic regurgitation is a complication of mechanical valve replacement. We describe a case of late valve dysfunction caused by an annular suture of excessive length obstructing the closure mechanism of a bileaflet prosthetic valve.We present this rare cause of valve dysfunction in an 80-year-old male patient who presented with haemolysis and dys...
متن کامل