Accuracy of two continuous glucose monitoring systems: a head-to-head comparison under clinical research centre and daily life conditions

نویسندگان

  • J Kropff
  • D Bruttomesso
  • W Doll
  • A Farret
  • S Galasso
  • Y M Luijf
  • J K Mader
  • J Place
  • F Boscari
  • T R Pieber
  • E Renard
  • J H DeVries
چکیده

AIMS To assess the accuracy and reliability of the two most widely used continuous glucose monitoring (CGM) systems. METHODS We studied the Dexcom®G4 Platinum (DG4P; Dexcom, San Diego, CA, USA) and Medtronic Paradigm Veo Enlite system (ENL; Medtronic, Northridge, CA, USA) CGM systems, in 24 patients with type 1 diabetes. The CGM systems were tested during 6-day home use and a nested 6-h clinical research centre (CRC) visit. During the CRC visit, frequent venous blood glucose samples were used as reference while patients received a meal with an increased insulin bolus to induce an aggravated postprandial glucose nadir. At home, patients performed at least six reference capillary blood measurements per day. A Wilcoxon signed-rank test was performed using all data points ≥15 min apart. RESULTS The overall mean absolute relative difference (MARD) value [standard deviation (s.d.)] measured at the CRC was 13.6 (11.0)% for the DG4P and 16.6 (13.5)% for the ENL [p < 0.0002, confidence interval of difference (CI Δ) 1.7-4.3%, n = 530]. The overall MARD assessed at home was 12.2 (12.0)% for the DG4P and 19.9 (20.5)% for the ENL (p < 0.0001, CI Δ = 5.8-8.7%, n = 839). During the CRC visit, the MARD in the hypoglycaemic range [≤3.9 mmol/l (70 mg/dl)], was 17.6 (12.2)% for the DG4P and 24.6 (18.8)% for the ENL (p = 0.005, CI Δ 3.1-10.7%, n = 117). Both sensors showed higher MARD values during hypoglycaemia than during euglycaemia [3.9-10 mmol/l (70-180 mg/dl)]: for the DG4P 17.6 versus 13.0% and for the ENL 24.6 versus 14.2%. CONCLUSIONS During circumstances of intended use, including both a CRC and home phase, the ENL was noticeably less accurate than the DG4P sensor. Both sensors showed lower accuracy in the hypoglycaemic range. The DG4P was less affected by this negative effect of hypoglycaemia on sensor accuracy than was the ENL.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

ارزیابی عملکرد دانه و برخی از صفات زراعی و مورفولوژیک در ژنوتیپ های گلرنگ بهاره در شرایط آبیاری و دیم

In order to assessment of grain yield and some of morphological characters in safflower genotypes under irrigated and rainfed conditions fifteen spring safflower genotypes were studied using two separate experiments based on randomized complete block design with three replications in 2010-2011 growing season at Miandoab Research Station, Iran. Analysis of variance for each conditions revealed s...

متن کامل

A Comparative Effectiveness Analysis of Three Continuous Glucose Monitors

OBJECTIVE To compare three continuous glucose monitoring (CGM) devices in subjects with type 1 diabetes under closed-loop blood glucose (BG) control. RESEARCH DESIGN AND METHODS Six subjects with type 1 diabetes (age 52 ± 14 years, diabetes duration 32 ± 14 years) each participated in two 51-h closed-loop BG control experiments in the hospital. Venous plasma glucose (PG) measurements (GlucoSc...

متن کامل

Evaluation of Six-Dimensional Cranial Target Positioning Accuracy in Two Different Immobilization Methods Using Exactrac System

Introduction: The aim of this study was to determine the accuracy of two different immobilization methods in patient positioning in cranial radiotherapy. The six-dimensional (6D) target localization accuracy of using a dedicated stereotactic mask was compared with that of a conventional head mask by the ExacTrac system. Material and Methods: ...

متن کامل

AP@home: a novel European approach to bring the artificial pancreas home.

The development of an artificial pancreas (AP) made huge strides from 2006 to 2008 and a large number of activities are going on in this area of research. Until now, most AP systems under development were tested only under highly controlled conditions. The aim of our project, funded by the European Union, is to develop an AP system to such a level that it can be studied under daily life conditi...

متن کامل

Development of A Head Docking Device for Linac-based Stereotactic Radiosurgery System to be Installed on Neptun 10 PC Linac

Introduction:  Stereotactic  radiosurgery  is  a  technique  for  the  treatment  of  intracranial  lesions  with  highly  collimated  x-ray  beam.  Linac-based  radiosurgery  is  currently  performed  by  two  techniques:  couch or pedestal mounted. In the first technique, a device is required to affix patient's head to a couch  and  moreover  to  translate  it  accurately.  Structure  of  suc...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 17  شماره 

صفحات  -

تاریخ انتشار 2015