Inaccuracy of seven popular sphygmomanometers for home measurement of blood pressure.

نویسندگان

  • E O'Brien
  • F Mee
  • N Atkins
  • K O'Malley
چکیده

Seven models, available commercially for the self-measurement of blood pressure, were subjected to a validation procedure in which three devices of each model were tested by observers who were trained to a high standard of accuracy. The models were the Omron HEM-400C, the Philips HP5308, the Healthcheck 'Cuffless' CX-5 060020, the Nissei Analogue Monitor, the Philips HP5306/B, the Systema Dr MI-150 and the Fortec Dr MI-100. The validation programme had a number of unique features which included assessment of interdevice variability before and after 1 month of home use, and a new form of analysis, which we term 'clinical', based on the likely influence of three grades of device inaccuracy on patient management. In the main validation phase, one device of each model was compared with simultaneous measurements made by two 'blinded' observers using a standard mercury sphygmomanometer (PyMaH Corporation, New Jersey, USA) in the same arm in 85 subjects with a wide range of blood pressures. Three models (the Healthcheck 'Cuffless' CX-5 060020, the Systema Dr MI-150 and the Fortec Dr MI-100) failed the interdevice variability tests and did not reach the main validation test. Two models (the Omron HEM-400C and the Philips HP5306/B) failed on the criteria set down by the American National Standard for Electronic or Automated Sphygmomanometers, as well as the 'clinical' criteria. The remaining two models (the Nissei Analogue Monitor and the Philips HP5308) were acceptable for the measurement of systolic blood pressure by both methods of analysis but failed in the 'clinical' analysis for diastolic blood pressure. The mercury sphygmomanometer was comfortably within the criteria for both methods of analysis.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

[Home blood pressure monitoring and control in a group of hypertensive patients].

This qualitative study was performed with 71 hypertensive patients, with the objectives to compare outpatient and home blood pressure monitoring (HBPM), to assess blood pressure control, and characterize white-coat hypertension. A nurse performed the outpatient blood pressure measurement. The home blood pressure monitoring was carried out over seven days. White-coat hypertension was quantified ...

متن کامل

Comparison of Various Blood Pressure Measurement Techniques with Standard Office Based Blood Pressure Mesurement

Introduction: In order to control Hypertension, it is necessary to understand the methods of blood pressure measurement. Recent guidelines emphasize on employing appropriate blood pressure measurement techniques and white coat hypertension. Definition and methods of blood pressure measurement have changed in the American guideline based on SPRINT trial (Systolic Blood Pressure Intervention Tria...

متن کامل

Proportion and characteristics of patients who measure their blood pressure at home: nationwide survey in Slovenia.

INTRODUCTION Home blood pressure monitoring has several advantages over blood pressure monitoring at a physician's office, and has become a useful instrument in the management of hypertension. OBJECTIVE To explore the rate and characteristics of patients who measure their blood pressure at home. METHODS A sample of 2,752 patients with diagnosis of essential arterial hypertension was selecte...

متن کامل

Demise of the mercury sphygmomanometer and the dawning of a new era in blood pressure measurement.

After a little more than a century of use, the conventional Riva-Rocci/Korotkoff technique of measuring blood pressure with a mercury sphygmomanometer and stethoscope, is now being relegated to the museum shelves. Affectionately attached though we may be to this clinical measurement, we must acknowledge that the technique is fraught with inaccuracy and that the age of technology has brought mor...

متن کامل

The Choice of Home Blood Pressure Monitoring Device Matters: Differences between Three Commercial Monitors Used in Finnish General Practice

The use of home measurement in the follow-up of antihypertensive treatment is rapidly increasing. Compared with office blood pressure (BP) measurement home blood pressure monitoring provides better reproducibility, freedom from white coat effect and observer bias [1,2]. Home BP measurement results correlate better to end organ damage than the office measurement and home BP control overcomes man...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of hypertension

دوره 8 7  شماره 

صفحات  -

تاریخ انتشار 1990