Standardization of a commercial (Boehringer Mannheim diagnostics) enzymic method for cholesterol.

نویسندگان

  • G R Warnick
  • M Spain
  • H Kloepfer
  • T M Volke
چکیده

The Laboratory Standardization Panel of the National Cholesterol Education Program recommends that cholesterol method accuracy ideally be within 3% of the true value determined by the Abell-Kendall Reference Method, a component of the National Reference System for Cholesterol. As one of the Abell-Kendall network laboratories established to facilitate cholesterol standardization, the approach we recommend for determining accuracy involves a comparison analysis on patients' specimens by the method in question and by the Abell-Kendall method. Use of fresh specimens precludes matrix interactions that may influence enzymic measurement. Using this approach, we assessed an enzymic method for cholesterol with two instruments (Boehringer Mannheim/Hitachi 717 and 737), with BMD reagent, controls, and calibrator. Fresh and frozen sera were analyzed with both instruments over three days. The Abell-Kendall method was used at the Northwest Lipid Research Center on frozen aliquots of the same sera. Both instruments demonstrated good agreement with the Reference Method, as determined by linear regression; overall bias averaged less than -2% for the Hitachi 717 and -1% for the Hitachi 737 at 2000 mg/L--i.e., within the accuracy recommendation. We observed a difference in bias for fresh and frozen specimens; with the Hitachi 717, fresh specimens exhibited -3% bias at 2000 mg/L, but there was virtually no bias of determinations of frozen specimens.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Assessment of renal function by serum creatinine and creatinine clearance: glomerular filtration rate estimated by four procedures.

We compared creatinine concentrations in serum and urine and creatinine clearances determined by two Jaffé (Beckman's "Astra," Boehringer Mannheim Diagnostics) and two enzymatic (Kodak, Boehringer Mannheim Diagnostics) methods. Serum creatinine and creatinine clearances determined by each method were also compared with the glomerular filtration rate as measured with use of sodium [125I]iothalam...

متن کامل

Four frequently used test systems for serum cholesterol evaluated by isotope dilution gas chromatography-mass spectrometry candidate reference method.

We evaluated the performance of four frequently used cholesterol test systems, using split-sample measurements with a panel of 79 patients' specimens and isotope dilution gas chromatography-mass spectrometry (ID GS-MS) as a comparison method. The test systems were from Beckman, a Boehringer Mannheim, Merck, and Johnson & Johnson Clinical Diagnostics, performed on the Synchron CX7, Hitachi 717, ...

متن کامل

Identification and treatment of hypomagnesemia and hypokalemia in patients receiving digoxin.

Hamsters are increasingly used for comparative study of cholesterol metabolism (1). To assess the validity of differential precipitation with Mg2 4-phosphotungstate for quantification of plasma high-density-lipoprotein cholesterol (HDLC), we did a method-comparison with ultracentrifugal flotation. The precipitating reagent, 0.55 mmolJL phosphotungstic acid and 25 mmol/L magnesium chloride, was ...

متن کامل

Enzymic serum cholesterol measurement with a basic autoanalyzer and Du Pont aca method.

I compared enzymic methods for total serum cholesterol as used with discrete (Du Pont aca and reagent packs) and continuous-flow (Boehringer Mannheim reagents and the Technicon AuotAnalyzer I) analysis of normal, icteric, and lipemic sera. The regression equation for 24 clear, non-icteric sera was: y (continuous flow) = 0.944x (aca) + 10.69; r = 0.971, Syx +/- 53.7 mg of cholesterol per liter. ...

متن کامل

Multicenter evaluation of Reflotron direct dry-chemistry assay of high-density lipoprotein cholesterol in venous and fingerstick specimens.

The Reflotron HDL Cholesterol test (Boehringer Mannheim GmbH) directly separates and analyzes high-density lipoprotein (HDL) cholesterol in plasma collected with EDTA in an integrated dry-reagent system suitable for alternative site testing of lipoproteins. We describe a multicenter evaluation of this test by two US and six European laboratories experienced in lipid analysis. Each laboratory co...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Clinical chemistry

دوره 35 3  شماره 

صفحات  -

تاریخ انتشار 1989