Automated quantification of creatine kinase MB isoenzyme in serum by radial partition immunoassay, with use of the Stratus analyzer.
نویسندگان
چکیده
We evaluated the analytical and clinical performances of a new radial partition immunoassay for measuring the mass concentration of creatine kinase (CK)-MB in serum. All pipetting, washes, incubations and data reduction were performed in 8 min by the Stratus (Dade) fluorometric analyzer. Within-assay and between-assay CVs were respectively 5.5% and 8.4% at 21 micrograms/L, and 4.2% and 3.4% at 48 micrograms/L. Assaying serial dilutions of serum samples with high CK-MB concentrations demonstrated excellent linearity. Results of the Stratus technique correlated well (n = 115, r = 0.98) with those of the Tandem-E CKMB II assay. There was no interference from hemolysis, bilirubin, rheumatoid factor, or added CK-MM (up to 3500 U/L); consequently, CK-MB can be determined in undiluted serum, even in the presence of high total CK activity. The mean CK-MB concentration in 105 blood donors was 1.9 (SD 1.3) micrograms/L. For seven myocardial infarction patients who received prompt fibrinolytic therapy, the mean CK-MB concentration was 4.5 (SD 1.8) micrograms/L at admission, and maximum concentrations, 119 (SD 94) micrograms/L, were recorded 16 h later. CK-MB returned to concentrations less than 10 micrograms/L within 72 h.
منابع مشابه
Creatine kinase MB measured by fluorometric enzyme immunoassay and immunochemiluminescence.
Creatine kinase MB (CK-MB) was measured in serum by a fluorometric enzyme immunoassay on the Stratus analyzer and by an immunochemiluminometric assay using the Ciba Corning Magic Lite System. Both methods were standardized against purified CK-MB, with Stratus underestimating by 20 percent and Magic Lite overestimating by 28 percent. The assays proved sensitive and linear; however, at a CK-MB co...
متن کاملImpact of antibody specificity and calibration material on the measure of agreement between methods for cardiac troponin I.
BACKGROUND Available assays for cardiac troponin I (cTnI) yield numerically different results. The aim of this study was to compare patient values obtained from four cTnI immunoassays. METHODS We studied the Stratus(R) II assay, the Opus(R) II assay, the Access(R) assay, and a research-only cTnI heterogeneous immunoassay that uses the Dade Behring aca(R) plus immunoassay system equipped with ...
متن کاملSemi-automated direct colorimetric measurement of creatine kinase isoenzyme MB activity after extraction from serum by use of a CK-MB-specific monoclonal antibody.
This semi-automated colorimetric assay for the MB isoenzyme of creatine kinase (EC 2.7.3.2) is based on a monoclonal antibody ("Conan-MB") specific for this isoenzyme and is a modification of a previously published method (Vaidya et al., Clin Chem 1986;32:657-63). A 0.64-cm bead coated with 2 to 3 micrograms of antibody is incubated with 100 microL of serum and 10 microL of 0.2 mol/L beta-merca...
متن کاملStratus automated creatine kinase-MB assay evaluated: identification and elimination of falsely increased results associated with a high-molecular-mass form of alkaline phosphatase.
We compared the performance of an automated assay of creatine kinase MB isoenzyme (CK-MB) mass (Stratus) with that of a CK-MB enzymatic assay routinely used at our institutions. Both of these assays use the same CK-MB-specific monoclonal antibody to immunocapture CK-MB, thus providing a direct means of comparing a mass assay with an activity assay. Routine CK-MB measurements for 206 samples wit...
متن کاملNew manual and automated method for determining activity of creatine kinase isoenzyme MB, by use of dithiothreitol: clinical applications.
The method is based on the selective activating capacity of dithiothreitol on creatine kinase isoenzyme MB, after isoenzyme MM is activated by glutathione. Isolated isoenzymes MM and MB of human and canine origin were assayed individually and in mixtures of known activities. When glutathione was present in the assay medium the activity of each isoenzyme could be measured individually, but gluta...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical chemistry
دوره 36 1 شماره
صفحات -
تاریخ انتشار 1990