Equations for calculating track parameters have been proposed, which invariably involve the track etch rate Vt and the bulk etch rate Vb. The present study measured Vb for the LR115 solid-state nuclear track detector using atomic force microscopy (AFM). The detectors were partially masked using rubber cement and then etched in 2:5 N NaOH solution at 60◦C for time periods ranging from 5 to 40 min. The rubber cement was then peeled o> and cross-sectional images of the LR115 detectors were obtained by AFM. Vb has been found to have di>erent values below and beyond the etching time of about 13:5 min, with the values of 0.0555 and 0:0875 m min−1, respectively. The increase in Vb with the etching time can be explained by a di>usion-etch model, in which the additional damage of the detector material is due to those etchant ions di>used into the detector over time. Now that Vb has been determined, this can be combined with the track etch rate Vt to calculate track parameters. c © 2002 Elsevier Science Ltd. All rights reserved.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

E ects of stirring on the bulk etch rate of LR 115 detector

The e ects of stirring on the bulk etch rate of LR 115 detector has been investigated. The surface pro6le measurement method using an instrument called Form Talysurf has been used to measure the thickness of the active layer of the LR 115 detectors. The etchant was 10% aqueous solution of NaOH maintained at 60◦C. The bulk etch rate under magnetic stirring has been found to be 6:65 ± 0:34 m h−1 ...

متن کامل

E ects of stirring on the bulk etch rate of CR-39 detector

It is well established that the bulk etch rates for solid state nuclear track detectors are a ected by the concentration and the temperature of the etchant. Recently, we found that the bulk etch rate for the LR 115 detector to be a ected by stirring during etching. In the present work, the e ects of stirring on the bulk etch rate of the CR-39 detector is investigated. One set of sample was etch...

متن کامل

Effect of Gamma Rays on The Bulk Etching Rate of LR-115 Detector

It is important to measure the bulk-etching rate for any solid state nuclear track detectors, because they have an application in different fields of sciences. The effect of gamma rays emitted from 60 Co source with different doses on LR-115 detector was studied. The bulk etching rate of the detectors was measured before and after irradiation using gamma source. From the results the values of t...

متن کامل

Fast and non-destructive determination of active-layer thickness of LR 115 SSNTD using a color commercial document scanner

Measurements using the LR 115 solid-state nuclear track detector (SSNTD) depend critically on the removed thickness of the active layer during etching, which cannot be controlled by the etching period alone. For example, the bulk etch rate depends significantly on the strength of stirring during etching. We propose here a fast, inexpensive and nondestructive method based on a color commercial d...

متن کامل

Non-destructive measurement of active-layer thickness of LR 115 SSNTD

LR 115 is a solid-state nuclear track detector (SSNTD) based on cellulose nitrate and has been commonly used for measurements of concentrations of radon gas and/or radon progeny. These measurements depend critically on the removed thickness of the active layer during etching. However, the thickness of removed layer calculated using the etching period does not necessarily provide a su6ciently ac...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2002