Albumin-based or albumin-linked calibrators cause a positive bias in serum proteins assayed by the biuret method.

نویسندگان

  • Vratislav Chromý
  • Lenka Sváchová
  • Lukás Novosád
  • Jirí Jarkovský
  • Pavel Sedlák
  • Petr Horák
  • Hana Dobrovolná
  • Barbora Hlavácová
چکیده

BACKGROUND Assay of total serum protein by the biuret method calibrated with albumin standards according to the reference method provides results with a positive bias approximately 3%-5% exceeding the total error of 3.4% allowable for total protein in serum analysis made by analysers using two-part reagents and short-term procedures. METHODS We used two types of two-part biuret reagents utilised in a short-term measurement in analysers with albumin or serum calibrators, in which protein was attested by the Kjeldahl method. RESULTS Tests with potentially interfering substances proved that serum blanking used in a short-term biuret procedure is not capable of sufficiently eliminating effects of serum interferents. A short-term blanking is evidently capable of suppressing only an absorbance caused by serum-present coloured and turbid interferents, but its capacity to transform them (oxidise, hydrolyse, saponify, etc.) to some other not-interfering substances is very low compared with a long-term blanking. Lipids and bilirubin are responsible for significant positive bias of total protein in normal serum samples (approximately 3%) and even a greater positive offset in lipaemic and icteric sera (approximately 5%). We verified that interference tests based on a normal serum spiked with endogenous lipids and bilirubin give quite false and misleading results in the biuret reaction. A pure albumin, not depending on its bovine/human origin, gives absorbance responding only to its copper complexes with protein with a biuret regent, while its absorbance with a serum also includes the absorbance of interferents present in serum. The simplest way to improve current short-term biuret procedures is the use of a human serum calibrator with total protein attested by the Kjeldahl method. A serum calibrator, behaving analogously to serum samples, compensates for a positive bias in most normal sera. Reagents with a greater concentration of active biuret components (copper and alkali, reference method included) seem to be unnecessarily aggressive to proteins and are responsible for a lower accuracy when used in short-term measurements. CONCLUSIONS Standard Reference Material 927c based on pure bovine albumin is still recommended and used as the primary standard for assays of total protein by colourimetric methods. The albumin calibrator is responsible for a positive bias of approximately 3%-5% in serum total protein assayed by the biuret reaction both in the reference and in current methods. Its substitution by a serum calibrator attested by the Kjeldahl method could solve this drawback. Clin Chem Lab Med 2009;47:91-101.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Four methods for determining total protein compared by using purified protein fractions from human serum.

The proportional bias of four methods frequently used for determining low concentrations of protein was evaluated with human serum protein fractions (Cohn Fractions II, III, IV, and V). Each fraction was assigned a protein concentration value as determined by the biuret method, calibrated with purified human serum albumin monomer. All four methods (Coomassie Brilliant Blue dye-binding, the meth...

متن کامل

Three turbidimetric methods for determining total protein compared.

We used human serum protein fractions to evaluate the sensitivity and bias of three turbidimetric methods for determining concentrations of proteins. Each fraction (Cohn Fractions II, III, IV, and V) was assigned a protein concentration value that was determined by the biuret method, which we calibrated with purified monomer of human serum albumin. All three turbidimetric methods (those involvi...

متن کامل

An improved method for determination of serum albumin and globulin.

A simple and reliable method has been developed for the analysis of serum proteins. Globulins are precipitated from serum with HCI-ethanol. The albumin remains in the supernatant fluid. The albumin fraction is then precipitated by alcoholic sodium acetate. The proteins are redissolved and determined colorimetrically following their reaction with the biuret reagent. Recovery and electrophoretic ...

متن کامل

The measurement of total serum proteins by the Biuret method.

The biuret reaction for proteins provides a simple and precise method for measuring serum proteins; Beer's law is obeyed to at least 10 g per dl. Several stable biuret reagents are available. Hemoglobin is the only important cause of interference which cannot be minimized by use of a sample blank. The mechanism of the biuret reaction is described and attention is drawn to the heterogeneity of t...

متن کامل

Relationship between serum levels of chromium and iron with transferrin and albumin levels as carrier proteins in diabetic patients with atherosclerosis

Backgrounds: Type 2 diabetes mellitus is one of the most common metabolic diseases in the world, resulting in complete or partial deficiency or resistance to insulin action. In this study, the association between serum levels of chromium and iron with serum levels of their carriers, transferrin and albumin in diabetic patients with atherosclerosis was investigated. Materials and methods: In ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Clinical chemistry and laboratory medicine

دوره 47 1  شماره 

صفحات  -

تاریخ انتشار 2009