Detection of bacteriuria by automated electrical impedance monitoring in a clinical microbiology laboratory.

نویسندگان

  • R Throm
  • S Specter
  • R Strauss
  • H Friedman
چکیده

An apparatus capable of rapidly detecting changes in electrical impedance was utilized for the continuous monitoring of bacterial growth in routine urine specimens in a clinical laboratory. In a trial study, 200 clinical specimens analyzed by the electrical impedance method resulted in an average detection time of 2.5 h for 41 clinically significant specimens, whereas conventional methods for bacterial isolation required overnight culture. Those specimens positive by the electrical impedance monitoring but negative by conventional bacteriological methods accounted for less than 2% of the total number of positive specimens, whereas electrical impedance-negative but conventional culture-positive specimens accounted for ca. 4%. Electrical impedance apparatus in clinical microbiology laboratories could provide rapid screening of clinical urine specimens as well as accurate detection of bacterial growth.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of the BACTOMETER microbial monitoring system with conventional methods for detection of microorganisms in urine specimens.

A rapid impedance-based screen for bacteriuria (less than 10(5) organisms per ml) was compared with a standard loop method, using 156 clinical urine specimens. Agreement between the methods was 96.8%.

متن کامل

Rapid detection of bacterial growth in blood samples by a continuous-monitoring electrical impedance apparatus.

A growth detection method utilizing an automated apparatus capable of rapidly detecting bacterial growth by measuring changes of electrical impedance in bacteriological medium was utilized with "mock" blood cultures containing various gram-negative and gram-positive bacteria. Measurement of changes of electrical impedance was found to ba as accurate and comparable for time of growth detection a...

متن کامل

Electrical impedance measurements: rapid method for detecting and monitoring microorganisms.

A conceptually simple and east-to-use technique is described that uses continuous impedance measurements for automated monitoring of microbial growth and metabolism. The method has been applied to a wide range of microorganisms. Optical clarity is not required. The sensitivity and reproducibility of the method are demonstrated. The mechanism whereby microbial growth alters the impedance of the ...

متن کامل

Analysis of the disagreement between automated bioluminescence-based and culture methods for detecting significant bacteriuria, with proposals for standardizing evaluations of bacteriuria detection methods.

A fully automated method for detecting significant bacteriuria is described which uses firefly luciferin and luciferase to detect bacterial ATP in urine. The automated method was calibrated and evaluated, using 308 urine specimens, against two reference culture methods. We obtained a specificity of 0.79 and sensitivity of 0.75 using a quantitative pour plate reference test and a specificity of ...

متن کامل

Evaluation of the leukocyte esterase and nitrite urine dipstick screening tests for detection of bacteriuria in women with suspected uncomplicated urinary tract infections.

A positive dipstick urinalysis (i.e., leukocyte esterase test and/or nitrite test) did not reliably detect significant bacteriuria in 479 ambulatory women with suspected uncomplicated urinary tract infection; 18.9% of the urine samples that demonstrated significant bacteriuria would have been rejected by the laboratory based on a negative urinalysis screen.

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of clinical microbiology

دوره 6 3  شماره 

صفحات  -

تاریخ انتشار 1977