Revisiting fetal dose during radiation therapy: evaluating treatment techniques and a custom shield

نویسندگان

  • Amir M. Owrangi
  • Donald A. Roberts
  • Elizabeth L. Covington
  • James A. Hayman
  • Kathryn M. Masi
  • Choonik Lee
  • Jean M. Moran
  • Joann I. Prisciandaro
چکیده

To create a comprehensive dataset of peripheral dose (PD) measurements from a new generation of linear accelerators with and without the presence of a newly designed fetal shield, PD measurements were performed to evaluate the effects of depth, field size, distance from the field edge, collimator angle, and beam modi-fiers for common treatment protocols and modalities. A custom fetal lead shield was designed and made for our department that allows external beam treatments from multiple angles while minimizing the need to adjust the shield during patient treatments. PD measurements were acquired for a comprehensive series of static fields on a stack of Solid Water. Additionally, PDs from various clinically relevant treatment scenarios for pregnant patients were measured using an anthropomorphic phantom that was abutted to a stack of Solid Water. As expected, the PD decreased as the distance from the field edge increased and the field size decreased. On aver-age, a PD reduction was observed when a 90° collimator rotation was applied and/or when the tertiary MLCs and jaws defined the field aperture. However, the effect of the collimator rotation (90° versus 0°) in PD reduction was not found to be clini-cally significant when the tertiary MLCs were used to define the field aperture. In the presence of both the MLCs and the fetal shield, the PD was reduced by 58% at a distance of 10 cm from the field edge. The newly designed fetal shield may effectively reduce fetal dose and is relatively easy to setup. Due to its design, we are able to use a broad range of treatment techniques and beam angles. We believe the acquired comprehensive PD dataset collected with and without the fetal shield will be useful for treatment teams to estimate fetal dose and help guide decisions on treat-ment techniques without the need to perform pretreatment phantom measurements.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Revisiting fetal dose during radiation therapy: evaluating treatment techniques and a custom shield [JACMP, 17(5), 2016]

This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. © 2017 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. TAB L E 7 The estimated fetal dos...

متن کامل

Design of a generally applicable abdominal shield for reducing fetal dose during radiotherapy of common malignancies in pregnant patients

Background: In most cancer cases, the treatment choice for a pregnant patient is radiotherapy. In these patients, the abdomen is usually not exposed therefore fetus exposure is due to peripheral dose (PD). The purpose of this study was to estimate the fetal dose (the maximum PD in each pregnancy stage) for modalities available and to fabricate and evaluate a generally applicable fetal ...

متن کامل

Designing a shield to reduce radiation dose during mammography: Dosimetric evaluation

Background: This paper presents a method to reduce radiation exposure during mammography by analysing the doses to ipsilateral and contralateral breasts and to adjacent organs by evaluating material-dependent shielding performance.  Materials and Methods: Six target-filter combinations (Mo-Mo, Mo-Rh, Rh-Rh, Rh-Mo, W-Mo, W-Rh) were tested by measuring the doses delivered to the breasts and adjac...

متن کامل

Evaluation of Breast Cancer Radiation Therapy Techniques in Outfield Organs of Rando Phantom with Thermoluminescence Dosimeter

Background: Given the importance of scattered and low doses in secondary cancer caused by radiation treatment, the point dose of critical organs, which were not subjected to radiation treatment in breast cancer radiotherapy, was measured.Objective: The purpose of this study is to evaluate the peripheral dose in two techniques of breast cancer radiotherapy with two energies. Methods: Eight diff...

متن کامل

Evaluation of Testis functioning during Pelvic Radiotherin the presence of testis lead shield

Introduction: Rectal and bladder cancer are the most common disease among young man. Although External Radiotherapy (EBRT) is one of the most effective treatment options introduced so far, the testis undesired radiation dose increases the risk of infertility and hormonal malfunctioning. The testicles are usually outside of the radiation field, but they can be exposed either to ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 17  شماره 

صفحات  -

تاریخ انتشار 2016