Validation of the Omron HEM-7320-LA, upper arm blood pressure monitor with Intelli Wrap Technology Cuff HEM-FL1 for self-measurement and clinic use according to the European Society of Hypertension International Protocol revision 2010 in the Mexican population
نویسندگان
چکیده
OBJECTIVE The aim of this study was to determine the accuracy of the Omron HEM-7320-LA with Intelli Wrap technology cuff HEM-FL1 for self-measurement and clinic blood pressure (BP) measurement according to the European Society of Hypertension International Protocol revision 2010. PARTICIPANTS AND METHODS The evaluation was performed in 39 individuals. The mean age of the participants was 47.9±14 years; systolic BP was 145.2±24.3 mmHg (range: 97-190), diastolic BP was 90.9±12.9 mmHg (range: 68-120), and arm circumference was 30.8±4 cm (range: 25-38.5). RESULTS The device successfully fulfilled the established criteria of the validation protocol. The device overestimated systolic BP by 0.6±5.7 mmHg and diastolic BP by 2.2±5.1 mmHg. The specially designed cuff HEM-FL1 to cover a broad range of arm circumferences and self-placement fulfilled the requirements of the International Protocol.
منابع مشابه
Validation of Omron RS8, RS6, and RS3 home blood pressure monitoring devices, in accordance with the European Society of Hypertension International Protocol revision 2010
BACKGROUND Allowing patients to measure their blood pressure at home is recognized as being of clinical value. However, it is not known how often these measurements are taken correctly. Blood pressure monitors for home use fall into two types based on the position of the cuff, ie, at the upper arm or the wrist. The latter is particularly convenient, as measurements can be taken fully clothed. T...
متن کاملValidation of four devices: Omron M6 Comfort, Omron HEM-7420, Withings BP-800, and Polygreen KP-7670 for home blood pressure measurement according to the European Society of Hypertension International Protocol
BACKGROUND Four oscillometric devices, including the Omron M6 Comfort, Omron HEM-7420, Withings BP-800, and Polygreen KP-7670, designed for self-blood pressure measurement (SBPM) were evaluated according to the European Society of Hypertension (ESH) International Protocol Revision 2010 in four separate studies. METHODS The four devices measure brachial blood pressure (BP) using the oscillomet...
متن کاملValidation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: the Omron M6 (HEM-7001-E) and the Omron R7 (HEM 637-IT).
BACKGROUND Two electronic devices for self-measurement of blood pressure - a brachial monitor, the Omron M6, and a wrist monitor, the Omron R7 - were evaluated in two separate studies according to the International Protocol of the European Society of Hypertension. DESIGN The International Validation Protocol is divided into two phases: the first phase is performed on 15 selected participants ...
متن کاملValidation of two automatic devices for the self-measurement of blood pressure according to the ANSI/AAMI/ISO81060-2:2009 guidelines: the Omron BP765 (HEM-7311-ZSA) and the Omron BP760N (HEM-7320-Z)
BACKGROUND Allowing patients to measure their blood pressure (BP) at home will be the standard for evaluating the disease state as the process of clinical diagnosis, and it is recognized as having great clinical utility. To measure BP as accurately as possible, innovative techniques have been incorporated into home BP measurement devices. OBJECTIVE The present study aimed to evaluate the perf...
متن کاملValidation of the YuWell YE690A upper-arm blood pressure monitor, for clinic use and self-measurement, according to the European Society of Hypertension International Protocol revision 2010.
OBJECTIVE The present study aimed to evaluate the accuracy of the automated oscillometric upper-arm blood pressure monitor YuWell YE690A for blood pressure measurement according to the International Protocol of the European Society of Hypertension revision 2010. PARTICIPANTS AND METHODS Systolic and diastolic blood pressures were measured sequentially in 33 adult Chinese (12 women, 44.2 years...
متن کامل