Monte Carlo calculation of beam quality correction for solid‐state detectors and phantom scatter correction at 137Cs energy

نویسندگان

  • T. Palani Selvam
  • Subhalaxmi Mishra
  • R.S. Vishwakarma
چکیده

Beam quality correction kQQo (r), which reflects the absorbed energy dependence of the detector, is calculated for solid state detector materials diamond, LiF, Li2B4O7 and Al2O3 for the 137Cs RTR brachytherapy source using the Monte Carlo-based EGSnrc code system. The study also includes calculation of detector-specific phantom scatter corrections kphant (r) for solid phantoms such as PMMA, polystyrene, RW1, solid water, virtual water and plastic water. Above corrections are calculated as a function of distance r along the transverse axis of the source. kQQo (r) is about unity for the Li2B4O7 detector. LiF detector shows a gradual decrease in kQQo (r) with r (decrease is about 2% over the distance range of 1 - 15 cm). Diamond detector shows a gradual increase in kQQo (r)with r (about 3% larger than unity at 15 cm). In the case of Al2O3 detector, kQQo (r)decreases with r steeply (about 14% over the distance range of 1 - 15 cm). The study shows that some solid state detectors demonstrate distance-dependent kphant (r)values, but the degree deviation from unity depends on the type of solid phantom and the detector.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Determination of the Energy Windows for the Triple Energy Window Scatter Correction Method in Gadolinium-159 Single Photon Emission Computed Tomography Using Monte Carlo Simulation

Introduction: In radionuclide imaging, object scatter is one of the major factors leading to image quality degradation. Therefore, the correction of scattered photons might have a great impact on improving the image quality. Regarding this, the present study aimed to determine the main and sub-energy windows for triple energy window (TEW) scatter correction method usin...

متن کامل

Monte Carlo Calculation of Beam Quality and Phantom Scatter Corrections for Lithium Formate Electron Paramagnetic Resonance Dosimeter for High-energy Brachytherapy Dosimetry

PURPOSE To investigate beam quality correction, KQQ0 (r) and phantom scatter correction, Kphan (r) for lithium formate dosimeter as a function of distance r along the transverse axis of the high-energy brachytherapy sources 60Co, 137Cs, 192Ir and 169Yb using the Monte Carlo-based EGSnrc code system. MATERIALS AND METHODS The brachytherapy sources investigated in this study are BEBIG High Dose...

متن کامل

Estimation of Dosimetric Parameters based on KNR and KNCSF Correction Factors for Small Field Radiation Therapy at 6 and 18 MV Linac Energies using Monte Carlo Simulation Methods

Background: Estimating dosimetric parameters for small fields under non-reference conditions leads to significant errors if done based on conventional protocols used for large fields in reference conditions. Hence, further correction factors have been introduced to take into account the influence of spectral quality changes when various detectors are used in non-reference conditions at differen...

متن کامل

Monte Carlo‐based beam quality and phantom scatter corrections for solid‐state detectors in 60Co and 192Ir brachytherapy dosimetry

Beam quality correction, kQQ0(r), for solid-state detectors diamond, LiF, Li2B4O7, Al2O3, and plastic scintillator are calculated as a function of distance, r, along the transverse axis of the 60Co and 192Ir brachytherapy sources using the Monte Carlo- based EGSnrc code system. This study also includes calculation of detector-specific phantom scatter correction, kphan(r), for solid phantoms suc...

متن کامل

The Effect of FEW Scatter Correction Method in In-111 Imaging

Introduction: In nuclear medicine imaging, detection of the scattered photons along with the primary photons is one of the major problems that can lead to a decrease in ddiagnostic accuracy. Therefore, use of a suitable scatter correction method can help to improve of the image quality and increase of diagnostic accuracy. The aim of this study is evaluation of five energy wind...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 15  شماره 

صفحات  -

تاریخ انتشار 2014