Endovascular beta-radiation to reduce restenosis after coronary balloon angioplasty: results of the beta energy restenosis trial (BERT).

نویسندگان

  • S B King
  • D O Williams
  • P Chougule
  • J L Klein
  • R Waksman
  • R Hilstead
  • J Macdonald
  • K Anderberg
  • I R Crocker
چکیده

BACKGROUND In the porcine overstretch injury model of restenosis, endovascular beta-radiation reduces neointima formation. To determine whether this therapy could be applied to patients with coronary artery disease, a special device was developed to allow delivery of 12 encapsulated 90Sr/Y sources, measuring a total of 30 mm, to various sites within the coronary arterial tree. This study was designed to evaluate the feasibility of the delivery of 12, 14, or 16 Gy at 2 mm after balloon angioplasty of stenoses of native coronary vessels. METHODS AND RESULTS Delivery of beta-radiation was attempted in 23 patients after successful balloon angioplasty. Source delivery was successful in 21 of the 23 patients (91%). There was no in-hospital or 30-day morbidity or mortality. Follow-up quantitative coronary arteriography in 20 patients demonstrated a late loss of 0.05 mm, a late loss index of 4%, and a restenosis rate of 15%. The use of the beta-emitter 90Sr/Y significantly reduced treatment time and operator exposure compared with previous trials with the gamma-emitter 192Ir. CONCLUSIONS In this study, the administration of endovascular beta-radiation after angioplasty was safe and feasible and substantially altered the postangioplasty late lumen loss, resulting in a lower-than-expected rate of restenosis. On the basis of these encouraging results, a multicenter, randomized trial with operators and patients blinded to treatment assignment is planned.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Endovascular b-Radiation to Reduce Restenosis After Coronary Balloon Angioplasty Results of the Beta Energy Restenosis Trial (BERT)

Background—In the porcine overstretch injury model of restenosis, endovascular b-radiation reduces neointima formation. To determine whether this therapy could be applied to patients with coronary artery disease, a special device was developed to allow delivery of 12 encapsulated Sr/Y sources, measuring a total of 30 mm, to various sites within the coronary arterial tree. This study was designe...

متن کامل

Dosimetry of 188Re and 186Re sources based on Monte Carlo calculations for endovascular brachytherapy after balloon angioplasty

Background: Recent pre-clinical and clinical studies indicate that irradiation in the dose range of 15 to 30 Gy can reduce rate of restenosis in patients who have undergone an angioplasty. The use of filled balloon with radioactive solution was proposed as one of the possible intravascular irradiation techniques. Materials and Methods: The Monte Carlo N-particle Transport Code (MCNP4b) was used...

متن کامل

Endovascular b-Radiation to Reduce Restenosis After Coronary Balloon Angioplasty

Background—In the porcine overstretch injury model of restenosis, endovascular b-radiation reduces neointima formation. To determine whether this therapy could be applied to patients with coronary artery disease, a special device was developed to allow delivery of 12 encapsulated Sr/Y sources, measuring a total of 30 mm, to various sites within the coronary arterial tree. This study was designe...

متن کامل

Effects of intracoronary beta-radiation therapy after coronary angioplasty: an intravascular ultrasound study.

BACKGROUND Endovascular radiation is emerging as a potential solution for the prevention and treatment of restenosis. Its effects on the morphology of unstented vessels cannot be determined by angiography and therefore require the use of intravascular ultrasound. METHODS AND RESULTS Through a 5F noncentered catheter for delivery of a 90Sr/Y source train, 12, 14, or 16 Gy at 2 mm was delivered...

متن کامل

Prevention of restenosis after angioplasty in small coronary arteries with probucol.

BACKGROUND Restenosis remains the major limitation of coronary angioplasty. Coronary stents have reduced the incidence of restenosis in selected patients with relatively large vessels. No strategies to date have demonstrated a beneficial effect in vessels < 3.0 mm in diameter. We have shown in the MultiVitamins and Probucol (MVP) Trial that probucol, a potent antioxidant, reduces restenosis aft...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Circulation

دوره 97 20  شماره 

صفحات  -

تاریخ انتشار 1998