Endotracheal cardiac output monitor.

نویسندگان

  • A W Wallace
  • A Salahieh
  • A Lawrence
  • K Spector
  • C Owens
  • D Alonso
چکیده

BACKGROUND The endotracheal cardiac output monitor (ECOM) is a new device that uses an endotracheal tube with multiple electrodes to measure cardiac output (CO). It measures the changes in electrical impedance caused by pulsatile blood flow in the aorta. The system was tested for safety and efficacy in 10 swine. METHODS Swine (60-80 kg) were chronically instrumented with a transit time flow probe on the ascending aorta and vascular occluders on the vena cava and pulmonary artery. After a minimum recovery of 4 days, the animals were anesthetized and intubated with an ECOM endotracheal tube. CO measurements from the ECOM system were compared to transit time flow probe measurements using linear regression and Bland-Altman analysis. Three different inotropic states were studied: (1) baseline; (2) increased (dobutamine); and (3) decreased (esmolol). CO was changed at each inotropic state by impeding left ventricular filling with the vena cava or pulmonary artery occluders. CO values between 0 and 15 l/min were studied. Pigs were studied for 24 h consecutively. RESULTS There was no deterioration of the impedance signal with time and no tracheal injury from the ECOM electrodes. There is a linear relationship between the ECOM and transit time flow probe CO between 0 and 15 l/min (slope = 0.94; intercept = 0.15 l/min; R2= 0.77). The mean difference between the two measures (bias) is 0.15 l/min and the SD is 1.34 l/min. The limits of agreement are -2.53 to 2.82 l/min. CONCLUSION Endotracheal CO monitor is a promising technology that needs further evaluation in clinical trials.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Non-invasive cardiac output monitor validation study in pediatric cardiac surgery patients.

STUDY OBJECTIVE To validate a non-invasive cardiac output monitor in pediatric cardiac surgery patients. DESIGN Prospective trial. SETTING Operating room. PATIENTS 20 pediatric cardiac surgery patients ASA physical status 3 and 4. INTERVENTIONS Aesculon noninvasive cardiac output monitor was used and compared to the cardiac output derived from the Fick equation. Oxygen consumption was m...

متن کامل

The effect of expiratory rib cage compression before endotracheal suctioning on the vital signs in patients under mechanical ventilation

BACKGROUND In patients undergoing mechanical ventilation, mucus production and secretion is high as a result of the endotracheal tube. Because endotracheal suction in these patients is essential, chest physiotherapy techniques such as expiratory rib cage compression before endotracheal suctioning can be used as a means to facilitate mobilizing and removing airway secretion and improving alveola...

متن کامل

Clinical evaluation of USCOM ultrasonic cardiac output monitor in cardiac surgical patients in intensive care unit.

BACKGROUND The USCOM ultrasonic cardiac output monitor (USCOM Pty Ltd, Coffs Harbour, NSW, Australia) is a non-invasive device that determines cardiac output by continuous-wave Doppler ultrasound. The aim of this study was to evaluate the accuracy of the USCOM device compared with the thermodilution technique in intensive care patients who had just undergone cardiac surgery. METHODS We conduc...

متن کامل

Esophageal Doppler: noninvasive cardiac output monitor.

In this article we describe the esophageal Doppler, a noninvasive, instantaneous cardiac output monitor. Its reliability has been demonstrated to be comparable to that of other current techniques used in the clinical arena to measure cardiac output. It helps guiding intravascular fluid resuscitation by quantifying the increase in flow in response to fluid challenges and by indicating the platea...

متن کامل

Why measure cardiac output?

Although cardiac output is a primary determinant of global O2 transport there are no absolute values that reflect circulatory adequacy, though very low values are of negative prognostic use. There is no agreement as to what constitutes a clinically relevant change in cardiac output. A recent clinical trial suggests that early goal-directed therapy aimed at increasing cardiac output improves sur...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Anesthesiology

دوره 92 1  شماره 

صفحات  -

تاریخ انتشار 2000