A Simple Temperature Compensation Method for Turbidity Sensor
نویسندگان
چکیده
Abstract. Turbidity sensor for water treatment applications are based on scattered light measurement. The electrooptical characteristic of light emitter and detector has a close relationship with the environmental temperature. Fluctuations in water temperature can potentially affect electronic components and cause output signal errors in turbidity sensor. Decrease or eliminate temperature error is necessary. With this background we have taken the temperature experiment of turbidity sensor developed by us. 11 different concentrations were measured, ranging from 0 to 100NTU. Each solution was cooled to 1°C, and then gradually heated to 40°C. The output signal of sensor increased according to the rise of temperature found from the experiment. In order to compensate the temperature errors, we developed a novel method to take temperature compensation by using software. The method carried by MCU which is used to compensate the measurement errors by soft programming is simple and effective; it can improve the error compensation precision on sensors greatly.
منابع مشابه
Research of Intelligent Turbidity Sensor
Turbidity is an important index to evaluate the water quality. Turbidity can reflect the effects of insoluble substances that contain bait and seston on water. Traditional methods of turbidity detection are complicated, they have low efficiency and poor reliability. To solve the turbidity detection problem in aquaculture, an intelligent optical turbidity sensor which is based on scattering theo...
متن کاملDevelopment of a Radial Pulse Tonometric (RPT) Sensor with a Temperature Compensation Mechanism
Several RPT sensors have been developed to acquire objective and quantitative pulse waves. These sensors offer improved performance with respect to pressure calibration, size and sensor deployment, but not temperature. Since most pressure sensors are sensitive to temperature, various temperature compensation techniques have been developed, but these techniques are largely inapplicable to RPT se...
متن کاملIn situ tryptophan-like fluorometers: assessing turbidity and temperature effects for freshwater applications.
Tryptophan-like fluorescence (TLF) is an indicator of human influence on water quality as TLF peaks are associated with the input of labile organic carbon (e.g. sewage or farm waste) and its microbial breakdown. Hence, real-time measurement of TLF could be particularly useful for monitoring water quality at a higher temporal resolution than available hitherto. However, current understanding of ...
متن کاملClosed-Loop Compensation of the Quadrature Error in MEMS Vibratory Gyroscopes
In this paper, a simple but effective method for compensation of the quadrature error in MEMS vibratory gyroscope is provided. The proposed method does not require any change in the sensor structure, or additional circuit in the feedback path. The mathematical relations of the proposed feedback readout system were analyzed and the proposed solution assures good rejection capabilities. Based on ...
متن کاملA Temperature Compensation Voltage Controlled Oscillator Using a Complementary to Absolute Temperature Voltage Reference
This paper presents a temperature compensation voltage controlled oscillator (VCO) based on Cross-Coupled pair and Colpitts structures which is suitable for military fields. Also, two inductors have been used for increasing the negative conductance. By using this method, start-up condition has been improved. Two varactors and a simple capacitor bank are applied for covering a wide tunning range...
متن کامل