Guidance to users of Nycomed Amersham and North American Scientific, Inc., I-125 interstitial sources: dosimetry and calibration changes: recommendations of the American Association of Physicists in Medicine Radiation Therapy Committee Ad Hoc Subcommittee on Low-Energy Seed Dosimetry.
نویسندگان
چکیده
Dose calculations to patients undergoing implantation of 125I interstitial brachytherapy sources are affected by two recent changes in low-energy seed dosimetry: (a) implantation of a new primary air-kerma strength standard at the National Institute of Standards and Technology (NIST) on 1 January 1999 and (b) publication of revised dose-rate distributions in AAPM's Task Group 43 Report. The guidance herein represents AAPM's recommendations for users of 125I interstitial seed products marketed prior to 1 January 1999 (Nycomed Amersham models 6711 and 6702 and North American Scientific, Inc. models 3631 A/S and 3631 A/M. Implementation of Task Group 43 (TG43) 125I dose calculations involves revising data stored in files of radiation treatment planning software and lowering the prescribed dose to be delivered to patients by as much as 15% to avoid modifying the dose actually delivered to patients. The magnitude of the dose prescription change depends on the dosimetry data used prior to TG43 and the implant geometry. Adapting to the revised NIST calibration standard requires the user to increase the dose-rate constant (or its equivalent by 11.5%) but does not require modification of the prescribed dose. Failure to correctly implement these modifications can result in 20% or even 30% errors.
منابع مشابه
TG-60 dosimetry parameters calculation for the β-emitter 153Sm brachytherapy source using MCNP
Background: The formalism recommended by Task Group 60 (TG-60) of the American Association of Physicists in Medicine (AAPM) is applicable for β sources. Radioactive biocompatible and biodegradable 153Sm glass seed without encapsulation is a β- emitter with a short half life and delivers a high dose rate to the tumor in the millimeter range. In this work the dosimetry parameters of the ...
متن کاملRecommendations of the American Association of Physicists in Medicine
In March 2004, the recommendations of the American Association of Physicists in Medicine sAAPMd on the interstitial brachytherapy dosimetry using I and Pd were reported in Medical Physics fTG-43 Update: Rivard et al., 31, 633–674 s2004dg. These recommendations include some minor changes in the dose-calculation formalism and a major update of the dosimetry parameters for eight widely used inters...
متن کاملUpdate of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculations.
Since publication of the American Association of Physicists in Medicine (AAPM) Task Group No. 43 Report in 1995 (TG-43), both the utilization of permanent source implantation and the number of low-energy interstitial brachytherapy source models commercially available have dramatically increased. In addition, the National Institute of Standards and Technology has introduced a new primary standar...
متن کاملComparison of dosimetry parameters of two commercially available Iodine brachytherapy seeds using Monte Carlo calculations
Background: Iodine brachytherapy sources with low photon energies have been widely used in treating cancerous tumors. Dosimetric parameters of brachytherapy sources should be determined according to AAPM TG-43U1 recommendations before clinical use. Monte Carlo codes are reliable tools in calculation of these parameters for brachytherapy sources. Materials and Methods: Dosimetric param...
متن کاملDosimetric characterization of a high dose rate 192I source for brachytherapy application using Monte Carlo simulation and benchmarking with thermoluminescent dosimetry
Background: The purpose of this project was to derive the brachytherapy dosimetric functions described by American Association of Physicists in Medicine (AAPM) TG-43 U1 based on high dose rate 192I sources. Materials and Methods: The method utilized included both simulation of the designed Polymethyl methacrylate (PMMA) phantom using the Monte Carlo of MCNP4C and benchmarking of the simulation ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Medical physics
دوره 26 4 شماره
صفحات -
تاریخ انتشار 1999