Efficacy of Heat Disinfection With the Aksys Personal Hemodialysis System.
نویسنده
چکیده
The first objective of this study was to validate the use of water at 85°C ± 5°C to achieve high-level disinfection of a clean-in-place extracorporeal dialysis circuit. The second objective was to demonstrate that applying this hot water method to the entire fluid pathways of a newly designed dialysis instrument, including an integral reverse-osmosis membrane, routinely allowed the production of back-filtered dialysate meeting the U.S. Pharmacopeia (USP) standard for the water for injection (WFI). In a first study, six dialyzers were inoculated with P. aeruginosa, incubated, and subjected to 75°C water for 30 minutes. No organisms could be recovered from the experimental dialyzers nor from three negative controls, but they were recovered from three positive controls. In a second study, the carbon tanks in the water inlet line and the dialysate tank were both inoculated with massive amounts of dialysis water-adapted Gram-negative organisms, followed by seven dialysis treatments using bovine blood, where the fifth and sixth procedures were separated by 2 idle days and the sixth and seventh procedures were separated by 3 idle days. In addition, the bovine blood used in each treatment was highly contaminated. In every case, the back-filtered dialysate met the WFI standard. From the above results we conclude that a hot water disinfection process is efficacious in reducing even grossly exaggerated contamination in the Aksys personal hemodialysis system to the point where it can produce water and dialysate that greatly exceed the Association for the Advancement of Medical Instrumentation (AAMI) standards and priming and rinse-back solution (back-filtered dialysate) that meets the USP standard for WFI.
منابع مشابه
نحوه بهکارگیری روشهای کنترل عفونت توسط کارکنان بخشهای همودیالیز
Background & Aim: Hemodialysis centers are highly susceptible to nasocomial infections due to application of hemodialysis machines, medical equipment, and invasive techniques and devices. This research was carried out in order to evaluate methods of infection control using by staff of hemodialysis centers.Methods & Materials: In this descriptive study, all hemodialysis units in hospitals of Teh...
متن کاملEffects of biofilm formation on haemodialysis monitor disinfection.
BACKGROUND Biofilms are composed of communities of micro-organisms adhering to essentially any surface. We evaluated whether biofilm formation in the hydraulic circuit of a purposely contaminated haemodialysis monitor would modify the efficacy of different disinfection modalities against bacteria and endotoxin concentrations. METHODS A water-borne Pseudomonas aeruginosa (109) suspension was r...
متن کاملInvestigating the effect of different disinfection methods on the efficiency, pressure drop and Figure of merit of medical masks
Background and aim:This study was conducted with the aim of determining the effect of different disinfection methods on the efficiency,pressure drop and quality of medical masks used by medical staff. Also,the filtration efficiency of medical masks was investigated in the present study Materials and methods:In the present study, 4 models of medical masks used by the treatment staff were sele...
متن کاملThe Effect of Educational Program on Physical Activity and Sleep in Hemodialysis patients: Applying the Self-Efficacy Theory
Background & objectives: Hemodialysis patients with a regular physical activity and proper sleep show better physical and mental health. As increasing self-efficacy of patients can increase these behaviors, this study was aimed to evaluate the effect of the self-efficacy theory on physical activity and sleep in hemodialysis patients. Methods: This research was an interventional study performed ...
متن کاملStudying the Effectiveness of Integrating Spiritual and Cognitive Counseling on Resiliency and Self-Efficacy of Hemodialysis Patients
Abstract Introduction: Kidney is one of the most distinguished organs of the body. Impairment in kidney function can cause serious harm to the person's physical and mental health. In the event of kidney failure, dialysis will perform kidney’s function, and the person will need dialysis for the rest of his or her life. Therefore, the aim of this study was to investigate the effectiveness of inte...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Home hemodialysis international. International Symposium on Daily Home Hemodialysis
دوره 1 1 شماره
صفحات -
تاریخ انتشار 1997