Comparing planning time, delivery time and plan quality for IMRT, RapidArc and tomotherapy
نویسندگان
چکیده
The purpose of this study is to examine plan quality, treatment planning time, and estimated treatment delivery time for 5- and 9-field sliding window IMRT, single and dual arc RapidArc, and tomotherapy. For four phantoms, 5- and 9-field IMRT, single and dual arc RapidArc and tomotherapy plans were created. Plans were evaluated based on the ability to meet dose-volume constraints, dose homogeneity index, radiation conformity index, planning time, estimated delivery time, integral dose, and volume receiving more than 2 and 5 Gy. For all of the phantoms, tomotherapy was able to meet the most optimization criteria during planning (50% for P1, 67% for P2, 0% for P3, and 50% for P4). RapidArc met less of the optimization criteria (25% for P1, 17% for P2, 0% for P3, and 0% for P4), while IMRT was never able to meet any of the constraints. In addition, tomotherapy plans were able to produce the most homogeneous dose. Tomotherapy plans had longer planning time, longer estimated treatment times, lower conformity index, and higher integral dose. Tomotherapy plans can produce plans of higher quality and have the capability to conform dose distributions better than IMRT or RapidArc in the axial plane, but exhibit increased dose superior and inferior to the target volume. RapidArc, however, is capable of producing better plans than IMRT for the test cases examined in this study.
منابع مشابه
Rotational IMRT techniques compared to fixed gantry IMRT and Tomotherapy: multi-institutional planning study for head-and-neck cases
BACKGROUND Recent developments enable to deliver rotational IMRT with standard C-arm gantry based linear accelerators. This upcoming treatment technique was benchmarked in a multi-center treatment planning study against static gantry IMRT and rotational IMRT based on a ring gantry for a complex parotid gland sparing head-and-neck technique. METHODS Treatment plans were created for 10 patients...
متن کاملDosimetric Evaluation of Intensity-Modulated Radiotherapy, Volumetric Modulated Arc Therapy, and Helical Tomotherapy for Hippocampal-Avoidance Whole Brain Radiotherapy
BACKGROUND Whole brain radiotherapy (WBRT) is a vital tool in radiation oncology and beyond, but it can result in adverse health effects such as neurocognitive decline. Hippocampal Avoidance WBRT (HA-WBRT) is a strategy that aims to mitigate the neuro-cognitive side effects of whole brain radiotherapy treatment by sparing the hippocampi while delivering the prescribed dose to the rest of the br...
متن کاملLung SBRT: dosimetric and delivery comparison of RapidArc, TomoTherapy, and IMRT
This study seeks to compare fixed-field intensity-modulated radiation therapy (FF IMRT), RapidArc (RA), and helical tomotherapy (HT) to discover the optimal treatment modality to deliver SBRT to the peripheral lung. Eight patients with peripheral primary lung cancer were reviewed. Plans were prescribed a dose of 48 Gy and optimized similarly with heterogeneity corrections. Plan quality was asse...
متن کاملImplementation of HybridArc treatment technique in preoperative radiotherapy of rectal cancer: dose patterns in target lesions and organs at risk as compared to helical Tomotherapy and RapidArc
PURPOSE HybridArc is a novel treatment technique blending aperture-enhanced optimized arcs with discrete IMRT-elements, allowing selection of arcs with a set of static IMRT-beams. This study compared this new technique to helical Tomotherapy, and RapidArc, in preoperative radiotherapy of rectal cancer. MATERIAL AND METHODS Twelve rectal cancer patients treated consecutively with Tomotherapy H...
متن کاملTarget splitting non-coplanar RapidArc radiation therapy for a diffuse sebaceous carcinoma of the scalp: a novel delivery technique
BACKGROUND AND PURPOSE To compare conventional lateral photon-electron, fixed-beam intensity modulated radiation therapy (IMRT), coplanar and non-coplanar RapidArc for the treatment of a diffuse sebaceous gland carcinoma of the scalp. METHODS Comprehensive dosimetry comparisons were performed among 3D-CRT, IMRT and various RapidArc plans. Target coverage, conformity index (CI), homogeneity in...
متن کامل