A Novel Prototype Neonatal Resuscitator That Controls Tidal Volume and Ventilation Rate: A Comparative Study of Mask Ventilation in a Newborn Manikin
نویسندگان
چکیده
The objective of this randomized controlled manikin trial was to examine tidal volume (VT) delivery and ventilation rate during mask positive pressure ventilation (PPV) with five different devices, including a volume-controlled prototype Next Step™ device for neonatal resuscitation. We hypothesized that VT and rate would be closest to target with the Next Step™. Twenty-five Neonatal Resuscitation Program providers provided mask PPV to a newborn manikin (simulated weight 1 kg) in a randomized order with a self-inflating bag (SIB), a disposable T-piece, a non-disposable T-piece, a stand-alone resuscitation system T-piece, and the Next Step™. All T-pieces used a peak inflation pressure of 20 cmH2O and a positive end-expiratory pressure of 5 cmH2O. The participants were instructed to deliver a 5 mL/kg VT (rate 40-60/min) for 1 min with each device and each of three test lungs with increasing compliance of 0.5, 1.0, and 2.0 mL/cmH2O. VT and ventilation rate were compared between devices and compliance levels (linear mixed model). All devices, except the Next Step™ delivered a too high VT, up to sixfold the target at the 2.0-mL/cmH2O compliance. The Next Step™ VT was 26% lower than the target in the low compliance. The ventilation rate was within target with the Next Step™ and SIB, and slightly lower with the T-pieces. In conclusion, routinely used newborn resuscitators over delivered VT, whereas the Next Step™ under delivered in the low compliant test lung. The SIB had higher VT and rate than the T-pieces. More research is needed on volume-controlled delivery room ventilation.
منابع مشابه
Neonatal resuscitation 2: an evaluation of manual ventilation devices and face masks.
BACKGROUND The key to successful neonatal resuscitation is effective ventilation. Little evidence exists to guide clinicians in their choice of manual ventilation device or face mask. The expiratory tidal volume measured at the mask (V(TE(mask))) is a good estimate of the tidal volume delivered during simulated neonatal resuscitation. AIM To compare the efficacy of (a) the Laerdal infant resu...
متن کاملNeonatal resuscitation 3: manometer use in a model of face mask ventilation.
BACKGROUND Adequate ventilation is the key to successful neonatal resuscitation. Positive pressure ventilation (PPV) is initiated with manual ventilation devices via face masks. These devices may be used with a manometer to measure airway pressures delivered. The expiratory tidal volume measured at the mask (V(TE(mask))) is a good estimate of the tidal volume delivered during simulated neonatal...
متن کاملLeak and obstruction with mask ventilation during simulated neonatal resuscitation.
OBJECTIVES To evaluate mask technique during simulated neonatal resuscitation and test the effectiveness of training in optimal mask handling. STUDY DESIGN Seventy participants(consultants, registrars and nurses) from neonatal units were asked to administer positive pressure ventilation at a flow of 8 l/min and a frequency of 40-60/min to a modified leak free, term newborn manikin (lung compl...
متن کاملEffects of decreasing inspiratory flow rate during simulated basic life support ventilation of a cardiac arrest patient on lung and stomach tidal volumes.
If the airway of a cardiac arrest patient is unprotected, basic life support with low rather than high inspiratory flow rates may reduce stomach inflation. Further, if the inspiratory flow rate is fixed such as with a resuscitator performance may improve; especially when used by less experienced rescuers. The purpose of the present study was to assess the effect of limited flow ventilation on r...
متن کاملMinute ventilation at different compression to ventilation ratios, different ventilation rates, and continuous chest compressions with asynchronous ventilation in a newborn manikin
BACKGROUND In newborn resuscitation the recommended rate of chest compressions should be 90 per minute and 30 ventilations should be delivered each minute, aiming at achieving a total of 120 events per minute. However, this recommendation is based on physiological plausibility and consensus rather than scientific evidence. With focus on minute ventilation (Mv), we aimed to compare today's stand...