Blood Metal Ion Thresholds to Identify Patients with Metal-on-Metal Hip Implants at Risk of Adverse Reactions to Metal Debris
نویسندگان
چکیده
BACKGROUND The authors of recent studies have reported newly devised implant-specific blood metal ion thresholds to predict adverse reactions to metal debris (ARMD) in patients who have undergone unilateral or bilateral metal-on-metal (MoM) hip arthroplasty. These thresholds were most effective for identifying patients at low risk of ARMD. We investigated whether these newly devised blood metal ion thresholds could effectively identify patients at risk of ARMD after MoM hip arthroplasty in an external cohort of patients. METHODS We performed a validation study involving 803 MoM hip arthroplasties (323 unilateral Birmingham Hip Resurfacing [BHR], 93 bilateral BHR, and 294 unilateral Corail-Pinnacle implants) performed in 710 patients at 3 European centers. All patients underwent whole-blood metal ion sampling, and were divided into 2 groups: those with ARMD (leading to revision or identified on imaging; n = 75) and those without ARMD (n = 635). Previously devised implant-specific blood metal ion thresholds (2.15 μg/L of cobalt for unilateral BHR; 5.5 μg/L for the maximum of either cobalt or chromium for bilateral BHR; and 3.57 μg/L of cobalt for unilateral Corail-Pinnacle implants) were applied to the validation cohort, and receiver operating characteristic curve analysis was used to establish the discriminatory characteristics of each threshold. RESULTS The area under the curve, sensitivity, specificity, and positive and negative predictive values for the ability of each implant-specific threshold to distinguish between patients with and without ARMD were, respectively, 89.4% (95% confidence interval [CI] = 82.8% to 96.0%), 78.9%, 86.7%, 44.1%, and 96.9% for unilateral BHR; 89.2% (CI = 81.3% to 97.1%), 70.6%, 86.8%, 54.5%, and 93.0% for bilateral BHR; and 76.9% (CI = 63.9% to 90.0%), 65.0%, 85.4%, 24.5%, and 97.1% for unilateral Corail-Pinnacle implants. Using the implant-specific thresholds, we missed 20 patients with ARMD (2.8% of the patients in this series). We missed more patients with ARMD when we used the fixed thresholds proposed by regulatory authorities: 35 (4.9%) when we used the U.K. threshold of 7 μg/L for both cobalt and chromium (p = 0.0003), 21 (3.0%) when we used the U.S. threshold of 3 μg/L for both cobalt and chromium (p = 1.0), and 46 (6.5%) when we used the U.S. threshold of 10 μg/L for both cobalt and chromium (p < 0.0001). CONCLUSIONS This external multicenter validation study confirmed that patients with blood metal ion levels below new implant-specific thresholds have a low risk of ARMD after MoM hip arthroplasty. Using these implant-specific thresholds, we missed fewer patients with ARMD compared with when the thresholds currently proposed by regulatory authorities were used. We therefore recommend using implant-specific blood metal ion thresholds when managing patients who have undergone MoM hip arthroplasty. LEVEL OF EVIDENCE Diagnostic Level III. See Instructions for Authors for a complete description of levels of evidence.
منابع مشابه
Can blood metal ion levels be used to identify patients with bilateral Birmingham Hip Resurfacings who are at risk of adverse reactions to metal debris?
AIMS We investigated whether blood metal ion levels could effectively identify patients with bilateral Birmingham Hip Resurfacing (BHR) implants who have adverse reactions to metal debris (ARMD). PATIENTS AND METHODS Metal ion levels in whole blood were measured in 185 patients with bilateral BHRs. Patients were divided into those with ARMD who either had undergone a revision for ARMD or had ...
متن کاملBlood metal ion levels are not a useful test for adverse reactions to metal debris
OBJECTIVES Alarm over the reported high failure rates for metal-on-metal (MoM) hip implants as well as their potential for locally aggressive Adverse Reactions to Metal Debris (ARMDs) has prompted government agencies, internationally, to recommend the monitoring of patients with MoM hip implants. Some have advised that a blood ion level >7 µg/L indicates potential for ARMDs. We report a systema...
متن کاملEarly failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: A consequence of excess wear.
Early failure associated with adverse reactions to metal debris is an emerging problem after hip resurfacing but the exact mechanism is unclear. We analysed our entire series of 660 metal-on-metal resurfacings (Articular Surface Replacement (ASR) and Birmingham Hip Resurfacing (BHR)) and large-bearing ASR total hip replacements, to establish associations with metal debris-related failures. Clin...
متن کاملMetal-on-Metal Hip Arthroplasty: A Review of Adverse Reactions and Patient Management
Recent alarming joint registry data highlighting increased revision rates has prompted further research into the area of metal-on-metal hip replacements and resurfacings. This review article examines the latest literature on the topic of adverse reactions to metal debris and summarises the most up-to-date guidelines on patient management. Adverse reactions to metal debris can cause significant ...
متن کاملThe clinical implications of elevated blood metal ion concentrations in asymptomatic patients with MoM hip resurfacings: a cohort study
OBJECTIVE To determine whether elevated blood cobalt (Co) concentrations are associated with early failure of metal-on-metal (MoM) hip resurfacings secondary to adverse reaction to metal debris (ARMD). DESIGN Cohort study. SETTING Single centre orthopaedic unit. PARTICIPANTS Following the identification of complications potentially related to metal wear debris, a blood metal ion screening...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 99 شماره
صفحات -