Use of venturi entrainment to deliver nasal high flow oxygen.indd
نویسندگان
چکیده
Objective: To evaluate the use of an adjustable venturi device with standard wall oxygen supply to deliver nasal high flow oxygen. Design: We set up a circuit using a standard 15 L/ min oxygen rotameter connected to a wall outlet, 2 m standard oxygen tubing, an adjustable venturi device, humidification chamber and nasal high flow circuit and cannulae. Delivered FiO2 and total gas flow rates were measured over a range of oxygen flow rates and venturi settings. The study was conducted in two parts a bench-top study to define the usable range and using a human subject to assess loaded performance of the circuit. Setting: Royal Adelaide Hospital Intensive Care Unit. Participants: One study author. Results: A clinically useful range of total flow rates (30-50 L/ min) and delivered FiO2 (0.4-0.6) was achieved using the circuit described. The variation in performance seen with loading of the circuit was clinically insignificant. We have calibrated this ‘Fisher and Paykel’ adjustable venturi device for use with the Fisher and Paykel heater chamber and circuit. Conclusion: We demonstrated that nasal high flow oxygen can be delivered in a clinically useful and predictable manner using a venturi entrainment device. From Royal Adelaide Hospital, North Terrace, Adelaide, Australia (Gerald L. Wong and Mark E. Finnis) Address for correspondence:
منابع مشابه
Compliance at night with low flow oxygen therapy: a comparison of nasal cannulae and Venturi face masks.
BACKGROUND The factors that influence nocturnal compliance among patients prescribed low flow oxygen therapy were determined and tolerance of nasal cannulae and Venturi face masks compared. METHODS Two studies were performed: (1) a prospective study of 99 hospitalised patients prescribed low flow oxygen therapy, 49 on nasal cannulae and 50 on Venturi face mask; and (2) a prospective study of ...
متن کاملNumerical Analysis of Water and Air in Venturi Tube to Produce Micro-Bubbles
Two-phase flow regimes are affected by conduit position, alignment, geometry, flow direction, physical characteristics, and flow rate of each phase as well as the heat flux toward the boundaries. Due to the importance of two-phase flow, numerous regimes have been identified. The first step in studying micro-bubble formation inside a venturi tube is to recognize the type of flow regimes. In this...
متن کاملThe value of high-flow nasal cannula oxygen therapy after extubation in patients with acute respiratory failure
OBJECTIVE To investigate the value of high-flow nasal cannula oxygen therapy after extubation in patients with acute respiratory failure. METHODS A single-center, prospective, randomized, controlled pilot trial was conducted between January 2013 and December 2014. Sixty enrolled patients were randomized immediately after extubation into either a high-flow nasal cannula group (n=30) or an air ...
متن کاملEnhanced aerodynamic reach of vapor and aerosol sampling for real-time mass spectrometric detection using Venturi-assisted entrainment and ionization.
Venturi-assisted ENTrainment and Ionization (VENTI) was developed, demonstrating efficient entrainment, collection, and transport of remotely sampled vapors, aerosols, and dust particulate for real-time mass spectrometry (MS) detection. Integrating the Venturi and Coandă effects at multiple locations generated flow and analyte transport from non-proximate locations and more importantly enhanced...
متن کاملNasal high-flow oxygen therapy system for improving sleep-related hypoventilation in chronic obstructive pulmonary disease: a case report
INTRODUCTION Sleep-related hypoventilation should be considered in patients with chronic obstructive pulmonary disease, because appropriate respiratory management during sleep is important for preventing elevation of PaCO2 levels. A nasal high-flow oxygen therapy system using a special nasal cannula can deliver suitably heated and humidified oxygen at up to 60 L/min. Since the oxygen concentrat...
متن کامل