A new device for assessing ankle dorsiflexion motion: reliability and validity.

نویسندگان

  • Jason Wilken
  • Smita Rao
  • Miriam Estin
  • Charles L Saltzman
  • H John Yack
چکیده

STUDY DESIGN Clinical measurement. OBJECTIVE To determine the validity and reliability of measures obtained using a custom-made device for assessing ankle dorsiflexion motion and stiffness. BACKGROUND Limited dorsiflexion has been implicated in the evolution of foot pain in a number of clinical populations. Assessment of ankle dorsiflexion range of motion (ROM) is, therefore, commonly performed as part of a foot and ankle examination. Conventional goniometric assessment methods have demonstrated limited intertester reliability, while alternative methods of measurements are generally more difficult to use. The Iowa ankle range of motion (IAROM) device was designed in an attempt to develop a simple, clinically relevant, and time- and cost-effective tool to measure ankle dorsiflexion range of motion and stiffness. METHODS Validity and intertester reliability of dorsiflexion range-of-motion measures using the IAROM device were assessed at 10, 15, 20, and 25 Nm of passively applied dorsiflexion torque, with both the knee extended and flexed approximately 20°. Stiffness (change in torque/change in dorsiflexion angle) values were determined using the angular change obtained between the 15- and 25-Nm torque levels. Convergent validity (n = 12) was assessed through comparison of ankle dorsiflexion angles measured simultaneously with the IAROM device and an optoelectronic motion analysis system. Intertester reliability (n = 17) was assessed by 2 testers who took measurements within the same day. RESULTS Validity testing demonstrated excellent agreement (intraclass correlation coefficient [ICC] values ranging from 0.95 to 0.98). Reliability testing demonstrated good to excellent intertester agreement (ICC values ranging from 0.90 to 0.95). The ICCs for ankle joint dorsiflexion stiffness were .71 and .85 for the knee in an extended and flexed position, respectively. CONCLUSION The IAROM device provides valid and reliable measurement of ankle dorsiflexion ROM. The IAROM device also allows calculation of stiffness by measuring ROM at multiple torque levels, although the reliability of the measurement is not optimal.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The concurrent validity and reliability of the Leg Motion system for measuring ankle dorsiflexion range of motion in older adults

BACKGROUND New reliable devices for range of motion (ROM) measures in older adults are necessary to improve knowledge about the functional capability in this population. Dorsiflexion ROM limitation is associated with ankle injuries, foot pain, lower limb disorders, loss of balance, gait control disorders and fall risk in older adults. The aim of the present study was to assess the validity and ...

متن کامل

Reliability and validity of a weight-bearing measure of ankle dorsiflexion range of motion.

PURPOSE To examine reliability and validity of the Lunge Test (LT) of dorsiflexion range of motion and determine the impact of different approaches to obtain a score on these parameters. METHODS Fifty-three patients with ankle injury/dysfunction provided initial assessment data for cross-sectional convergent and known-groups validity analysis with the Pearson coefficient (r) and paired t-test...

متن کامل

مقایسه زمان آغاز فعالیت عضلات منتخب مچ پا در ورزشکاران با و بدون محدودیت دامنه حرکتی دورسی‌فلکشن، حین پرش_فرود تک پا

Objective: Limitation of ankle dorsiflexion range of motion as the most important risk factors for ankle sprain injury can cause disorder in neuromuscular control of muscles around the ankle. The aim of this study was to compare the onset of the activity of selected muscles around the ankle joint of those athletes with ankle dorsiflexion range of motion limitations with healthy Athletes during ...

متن کامل

Comparison between the Short Term and Durability Effects of Proprioceptive Neuromuscular Facilitation Exercise and Instrument Assisted soft Tissue Mobilization Technique on Ankle-Dorsiflexion Range of Motion

Introduction: Reduced lower extremity range of motion and movement leads to a decrease in movement efficiency; accordingly, it is vital to maintain the range of motion to prevent damage and improve functions. The purpose of this study was to compare the short term and durability effects of Instrument Assisted Soft Tissue Mobilization (IASTM) and Proprioceptive Neuromuscular Facilitation (PNF) t...

متن کامل

The Effect of Thera-Band Resistive Exercises on Pain, Dynamic Balance, and Function of Amateur Teenage Basketball Players

 Aims and background: With respect to the prevalence of ankle injuries due to the decrease in range of motion of dorsiflexion, and to the lack of proper function in foot and/or lack of balance among basketball players, utilizing exercise plans is of great importance. Therefore, the present study aims to investigate the effect of eight weeks of thera-band exercises on the range of motion on pain...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • The Journal of orthopaedic and sports physical therapy

دوره 41 4  شماره 

صفحات  -

تاریخ انتشار 2011