Accuracy quantification of a deformable image registration tool applied in a clinical setting
نویسندگان
چکیده
The purpose of this study was to test the accuracy of a commercially available deformable image registration tool in a clinical situation. In addition, to demonstrate a method to evaluate the resulting transformation of such a tool to a reference defined by multiple experts. For 16 patients (seven head and neck, four thoracic, five abdominal), 30-50 anatomical landmarks were defined on recognizable spots of a planning CT and a corresponding fraction CT. A commercially available deformable image registration tool, Velocity AI, was used to align all fraction CTs with the respective planning CTs. The registration accuracy was quantified by means of the target registration error in respect to expert-defined landmarks, considering the interobserver variation of five observers. The interobserver uncertainty of the landmark definition in our data sets is found to be 1.2 ± 1.1 mm. In general the deformable image registration tool decreases the extent of observable misalignments from 4-8 mm to 1-4 mm for nearly 50% of the landmarks (to 77% in sum). Only small differences are observed in the alignment quality of scans with different tumor location. Smallest residual deviations were achieved in scans of the head and neck region (79%, ≤ 4 mm) and the thoracic cases (79%, ≤ 4 mm), followed by the abdominal cases (59%, ≤ 4 mm). No difference is observed in the alignment quality of different tissue types (bony vs. soft tissue). The investigated commercially available deformable image registration tool is capable of reducing a mean target registration error to a level that is clinically acceptable for the evaluation of retreatment plans and replanning in case of gross tumor change during treatment. Yet, since the alignment quality needs to be improved further, the individual result of the deformable image registration tool has still to be judged by the physician prior to application.
منابع مشابه
A Study on Robustness of Various Deformable Image Registration Algorithms on Image Reconstruction Using 4DCT Thoracic Images
Background: Medical image interpolation is recently introduced as a helpful tool to obtain further information via initial available images taken by tomography systems. To do this, deformable image registration algorithms are mainly utilized to perform image interpolation using tomography images.Materials and Methods: In this work, 4DCT thoracic images of five real patients provided by DI...
متن کاملEvaluation of deformable image registration in HDR gynecological brachytherapy
Introduction: In brachytherapy, as in external radiotherapy, image-guidance plays an important role. For GYN treatments it is standard to acquire at least CT images and preferably MR images prior to each treatment and to calculate the dose of the day on each set of images. Then, the dose to the target and to the organs at risk (OAR) is calculated with worst case scenario from I...
متن کاملEvaluation of the tool “Reg Refine” for user‐guided deformable image registration
"Reg Refine" is a tool available in the MIM Maestro v6.4.5 platform (www.mim-software.com) that allows the user to actively participate in the deformable image registration process. The purpose of this work was to evaluate the efficacy of this tool and investigate strategies for how to apply it effectively. This was done by performing DIR on two publicly available ground-truth models, the Pixel...
متن کاملA Novel Subsampling Method for 3D Multimodality Medical Image Registration Based on Mutual Information
Mutual information (MI) is a widely used similarity metric for multimodality image registration. However, it involves an extremely high computational time especially when it is applied to volume images. Moreover, its robustness is affected by existence of local maxima. The multi-resolution pyramid approaches have been proposed to speed up the registration process and increase the accuracy of th...
متن کاملA measure to evaluate deformable registration fields in clinical settings
Deformable registration has migrated from a research topic to a widely used clinical tool that can improve radiotherapeutic treatment accuracy by tracking anatomical changes. Although various mathematical formulations have been reported in the literature and implemented in commercial software, we lack a straightforward method to verify a given solution in routine clinical use. We propose a metr...
متن کامل