Early detection of negative BACTEC MGIT 960 cultures by PCR-reverse cross-blot hybridization assay.
نویسندگان
چکیده
We evaluated the efficacy of a PCR-reverse cross-blot hybridization assay, a test which permits identification of mycobacteria by means of species-specific probes and a Mycobacterium-specific probe, for early detection of negative BACTEC MGIT 960 mycobacterial cultures. Aliquots of 549 cultures were collected 7 days after the culture media were inoculated with various clinical specimens and tested with the molecular assay. PCR results were compared to those obtained at the end times with the BACTEC MGIT 960 system. Of the 549 specimens analyzed, 484 were found to be negative and 64 were found positive by both methods; one specimen, found to be positive by the BACTEC MGIT 960 system, was identified as negative by the molecular assay. In view of its high negative predictive value (99.8%), the PCR-reverse cross-blot hybridization assay appears to be a valid tool for early detection of negative BACTEC MGIT 960 cultures.
منابع مشابه
Growth detection failures by the nonradiometric Bactec MGIT 960 mycobacterial culture system.
Mycobacterial growth in liquid culture can go undetected by automated, nonradiometric growth detection systems. In our laboratory, instrument-negative tubes from the Bactec MGIT 960 system are inspected visually for clumps suggestive of mycobacterial growth, which (if present) are examined by acid-fast smear analysis. A 3-year review demonstrated that ∼1% of instrument-negative MGIT cultures co...
متن کاملComparison of the BACTEC MGIT 960 and ESP culture system II for growth and detection of mycobacteria.
The performances of two continuously monitoring mycobacterial culture systems-ESP Culture System II (ESP II; Trek Diagnostics, Inc. , Westlake, Ohio) and BACTEC MGIT 960 (BD Biosciences, Sparks, Md. )-were compared. In addition to both liquid media, all specimens were plated onto Middlebrook 7H11/7H11 selective agar. A total of 3, 151 specimens of all types (56.3% were respiratory specimens) we...
متن کامل[A study on inoculum density and reproducibility of drug susceptibility testing by BACTEC MGIT 960].
OBJECTIVE The BACTEC MGIT 960 drug susceptibility system (MGIT AST) has been recently introduced in Japan. The issue of discordant MGIT results compared with the conventionally used Ogawa method has been raised. It has been speculated that discordant results might be due to MGIT inoculum density since there is no standardization step other than dilution of growth for tubes beyond 2 days after M...
متن کاملComparative evaluation of BACTEC MGIT 960 with BACTEC 9000 MB and LJ for isolation of mycobacteria in The Gambia.
BACKGROUND The BACTEC MGIT 960 was evaluated and compared with BACTEC 9000 MB and Lowenstein-Jensen medium for recovery rate of mycobacteria, time to detection, and contamination rate. METHODOLOGY 147 sputum samples obtained from patients with suspicion of tuberculosis were processed and inoculated into BACTEC MGIT 960, BACTEC 9000 MB and Lowenstein-Jensen medium using standardized procedures...
متن کاملEvaluation of the BACTEC MGIT 960 and the MB/BacT systems for recovery of mycobacteria from clinical specimens and for species identification by DNA AccuProbe.
A total of 120 mycobacterial isolates were recovered from 1,068 clinical specimens. Of these, 82.5% were in MGIT 960, 83.3% were in MB/BacT, 80% were in BACTEC 460, and 70% were on Löwenstein-Jensen medium. Mean times to detection of Mycobacterium tuberculosis (n = 96) were significantly shorter with MGIT 960 (12.6 days, P = 0.003) and BACTEC 460 (11.8 days, P < 0.001) than with MB/BacT (15.9 d...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of clinical microbiology
دوره 40 9 شماره
صفحات -
تاریخ انتشار 2002