Complex conjugate resolved heterodyne swept source optical coherence tomography using coherence revival
نویسندگان
چکیده
We describe a simple and low-cost technique for resolving the complex conjugate ambiguity in Fourier domain optical coherence tomography (OCT) that is applicable to many swept source OCT (SSOCT) systems. First, we review the principles of coherence revival, wherein an interferometer illuminated by an external cavity tunable laser (ECTL) exhibits interference fringes when the two arms of the interferometer are mismatched by an integer multiple of the laser cavity length. Second, we report observations that the spectral interferogram obtained from SSOCT systems employing certain ECTLs are automatically phase modulated when the arm lengths are mismatched this way. This phase modulation results in a frequency-shifted interferogram, effectively creating an extended-depth heterodyne SSOCT system without the use of acousto-optic or electro-optic modulators. We suggest that this phase modulation may be caused by the ECTL cavity optical pathlength varying slightly over the laser sweep, and support this hypothesis with numerical simulations. We also report on the successful implementation of this technique with two commercial swept source lasers operating at 840nm and 1040nm, with sweep rates of 8kHz and 100kHz respectively. The extended imaging depth afforded by this technique was demonstrated by measuring the sensitivity fall-off profiles of each laser with matched and mismatched interferometer arms. The feasibility of this technique for clinical systems is demonstrated by imaging the ocular anterior segments of healthy human volunteers.
منابع مشابه
Complete complex conjugate resolved heterodyne swept source optical coherence tomography using a dispersive optical delay line: erratum
We correct an error in our previous paper [Biomed. Opt. Express 2, 1218 (2011)] which led to an erroneous conclusion that a dispersive optical delay line (DODL) used in a swept source optical coherence tomography (SSOCT) system generated a pure phase modulation allowing for complex conjugate artifact removal in Fourier domain OCT via optical heterodyning. We now understand that an alternate phe...
متن کاملComplete complex conjugate resolved heterodyne swept-source optical coherence tomography using a dispersive optical delay line
Swept-source optical coherence tomography (SSOCT) provides a substantial sensitivity advantage over its time-domain counterpart, but suffers from a reduced imaging depth range due to sensitivity falloff and complex conjugate ambiguity. Heterodyne complex conjugate-resolved SSOCT (HCCR-SSOCT) has been previously demonstrated as a technique to completely resolve the complex conjugate ambiguity, e...
متن کاملInstantaneous complex conjugate resolved spectral domain and swept-source OCT using 3x3 fiber couplers.
We report that the complex conjugate artifact in Fourier domain optical coherence tomography approaches (including spectral domain and swept source OCT) may be resolved by the use of novel interferometer designs based on 3x3 and higher order fiber couplers. Interferometers built from NxN (N>2) truly fused fiber couplers provide simultaneous access to non-complementary phase components of the co...
متن کاملMaster/slave interferometry - ideal tool for coherence revival swept source optical coherence tomography.
In this paper, we demonstrate that the master slave (MS) interferometry method can significantly simplify the practice of coherence revival swept source optical coherence tomography (OCT) technique. Previous implementations of the coherence revival technique required considerable resources on dispersion compensation and data resampling. The total tolerance of the MS method to nonlinear tuning, ...
متن کاملImaging limbal and scleral vasculature using Swept Source Optical Coherence Tomography.
We demonstrate an application of high-speed swept source optical coherence tomography for vessel visualization in the anterior segment of the human eye. The human corneo-scleral junction and sclera was imaged in vivo. Imaging was performed using a swept source OCT system operating at the 1050nm wavelength range and 100kHz A-scan rate. High imaging speed enables the generation of 3D depth-resolv...
متن کامل