Dose-to-medium vs. dose-to-water: Dosimetric evaluation of dose reporting modes in Acuros XB for prostate, lung and breast cancer

نویسندگان

  • Suresh Rana
  • Shyam Pokharel
چکیده

Purpose: Acuros XB (AXB) dose calculation algorithm is available for external beam photon dose calculations in Eclipse treatment planning system (TPS). The AXB can report the absorbed dose in two modes: dose-to-water (Dw) and dose-to-medium (Dm). The main purpose of this study was to compare the dosimetric results of the AXB_Dm with that of AXB_Dw on real patient treatment plans. Methods: Four groups of patients (prostate cancer, stereotactic body radiation therapy (SBRT) lung cancer, left breast cancer, and right breast cancer) were selected for this study, and each group consisted of 5 cases. The treatment plans of all cases were generated in the Eclipse TPS. For each case, treatment plans were computed using AXB_Dw and AXB_Dm for identical beam arrangements. Dosimetric evaluation was done by comparing various dosimetric parameters in the AXB_Dw plans with that of AXB_Dm plans for the corresponding patient case. Results: For the prostate cancer, the mean planning target volume (PTV) dose in the AXB_Dw plans was higher by up to 1.0%, but the mean PTV dose was within ±0.3% for the SBRT lung cancer. The analysis of organs at risk (OAR) results in the prostate cancer showed that AXB_Dw plans consistently produced higher values for the bladder and femoral heads but not for the rectum. In the case of SBRT lung cancer, a clear trend was seen for the heart mean dose and spinal cord maximum dose, with AXB_Dw plans producing higher values than the AXB_Dm plans. However, the difference in the lung doses between the AXB_Dm and AXB_Dw plans did not always produce a clear trend, with difference ranged from -1.4% to 2.9%. For both the left and right breast cancer, the AXB_Dm plans produced higher maximum dose to the PTV for all cases. The evaluation of the maximum dose to the skin showed higher values in the AXB_Dm plans for all 5 left breast cancer cases, whereas only 2 cases had higher maximum dose to the skin in the AXB_Dm plans for the right breast cancer. Conclusion: The preliminary dosimetric results from our clinical study showed that the selection of either Dm or Dw in AXB is less likely to produce significant dosimetric differences in the clinical environment. However, the difference between the AXB_Dm and AXB_Dw calculations depends on the disease site, and even for the same type of disease (e.g., lung cancer), the results are patient specific.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

On the dosimetric impact of inhomogeneity management in the Acuros XB algorithm for breast treatment

BACKGROUND A new algorithm for photon dose calculation, Acuros XB, has been recently introduced in the Eclipse, Varian treatment planning system, allowing, similarly to the classic Monte Carlo methods, for accurate modelling of dose deposition in media. Aim of the present study was the assessment of its behaviour in clinical cases. METHODS Datasets from ten breast patients scanned under diffe...

متن کامل

Clinical implementation and evaluation of the Acuros dose calculation algorithm

PURPOSE The main aim of this study is to validate the Acuros XB dose calculation algorithm for a Varian Clinac iX linac in our clinics, and subsequently compare it with the wildely used AAA algorithm. METHODS AND MATERIALS The source models for both Acuros XB and AAA were configured by importing the same measured beam data into Eclipse treatment planning system. Both algorithms were validated...

متن کامل

Dosimetric evaluation of Acuros XB Advanced Dose Calculation algorithm in heterogeneous media

BACKGROUND A study was realised to evaluate and determine relative figures of merit of a new algorithm for photon dose calculation when applied to inhomogeneous media. METHODS The new Acuros XB algorithm implemented in the Varian Eclipse treatment planning system was compared against a Monte Carlo method (VMC++), and the Analytical Anisotropic Algorithm (AAA). The study was carried out in vir...

متن کامل

Comparison of forward intensity modulated radiation therapy and two field tangential wedged beam techniques in radiotherapy of breast cancer: dosimetric and radiobiologic study

Introduction: This study aims to compare the dose delivery parameters and radiobiologic model prediction of Tumor Control Probability (TCP) of the FIMRT technique compared with rectangular two tangential wedged fields (2FW) in the treatment of breast cancer. Furthermore, reduction of the heart and lung dose in the FIMRT was studied via dosimetric parameters and radiobiologic mo...

متن کامل

Dosimetric accuracy of the Acuros XB and Anisotropic analytical algorithm near interface of the different density media for the small fields of a 6- MV flattening-filter-free beam

Background: This study was conducted to assess the accuracy of dose calculation near the air-phantom interface of a heterogeneous phantom for Acuros XB (AXB) and Anisotropic Analytical Algorithm (AAA) algorithm of a           6-MV flattening-filter-free beam, compared with film measurements. Materials and Methods: A phantom included air gap was ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2014