Impact of 3.0 T Cardiac MR Imaging Using Dual-Source Parallel Radiofrequency Transmission with Patient-Adaptive B1 Shimming

نویسندگان

  • Haipeng Jia
  • Cuiyan Wang
  • Guangbin Wang
  • Lei Qu
  • Weibo Chen
  • Queenie Chan
  • Bin Zhao
چکیده

OBJECTIVES To prospectively evaluate the impact of 3.0 T Cardiac MR imaging using dual-source parallel radiofrequency (RF) transmission with patient-adaptive B1 shimming compared with single-source RF transmission in the RF homogeneity, image contrast and image quality. METHODS The study was approved by the local institutional review board, and all subjects provided written informed consent. Fourteen healthy volunteers were examined at 3.0 T MR, with both the conventional single-source and the new dual-source RF transmission. B1 calibrations (RF shimming) of the heart region were performed to acquire a percent of the prescribed flip angle (FA) of B1 maps, which were used for quantitative assessment of RF homogeneity. Contrast ratios (CRs) between ventricular blood pool and septum were calculated on balanced-turbo field echo (B-TFE) cine images. The off-resonance artifacts of cine images were blindly assessed by two radiologists according to a 4-point grading-scale. RESULTS A significantly lower mean coefficients of variance of the achieved FA with dual-source revealed better RF homogeneity compared to single-source (P = 0.0094). Dual-source RF shimming significantly increased the CRs (P<0.05) and reduced the off-resonance artifacts of B-TFE cine images (P<0.05). Inter-observer agreement for the off-resonance artifacts of B-TFE cine images was good to excellent (k >0.65). CONCLUSIONS Dual-source parallel RF transmission significantly improves the RF homogeneity, increases image contrast and reduces image artifacts of cardiac B-TFE images compared to single-source mode. This may be of value in reducing the observer-dependence of cardiac MR images and enhancing diagnostic confidence for clinical practice using CMR at 3.0 T.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The impact of dual-source parallel radiofrequency transmission with patient-adaptive shimming on the cardiac magnetic resonance in children at 3.0 T

The cardiac magnetic resonance (CMR) of children at 3.0 T presents a unique set of technical challenges because of their small cardiac anatomical structures, fast heart rates, and the limited ability to keep motionless and hold breathe, which could cause problems associated with field inhomogeneity and degrade the image quality. The aim of our study was to evaluate the effect of dual-source par...

متن کامل

Clinical liver MRI at 3.0 Tesla using parallel RF transmission with patient-adaptive B1 shimming

Introduction: The clinical implementation of high field MRI systems has introduced new challenges for body imaging with respect to B1 field nonuniformities, which can result in the generation of standing waves and signal loss particularly in abdominal imaging. Several approaches have been proposed to achieve a more homogeneous excitation and to avoid a degradation of image quality at higher mag...

متن کامل

Evaluation of a Subject specific dual-transmit approach for improving B1 field homogeneity in cardiovascular magnetic resonance at 3T

BACKGROUND Radiofrequency (RF) shading artifacts degrade image quality while performing cardiovascular magnetic resonance (CMR) at higher field strengths. In this article, we sought to evaluate the effect of local RF (B1 field) shimming by using a dual-source-transmit RF system for cardiac cine imaging and to systematically evaluate the effect of subject body type on the B1 field with and witho...

متن کامل

Extended RF shimming: Sequence‐level parallel transmission optimization applied to steady‐state free precession MRI of the heart

Cardiac magnetic resonance imaging (MRI) at high field presents challenges because of the high specific absorption rate and significant transmit field (B1+ ) inhomogeneities. Parallel transmission MRI offers the ability to correct for both issues at the level of individual radiofrequency (RF) pulses, but must operate within strict hardware and safety constraints. The constraints are themselves ...

متن کامل

SAR optimised local B1 + shimming for cardiac imaging at 3T – a multi-model study

Target Audience: This work will benefit those interested in using parallel transmission MRI to reduce SAR in cardiac MR Purpose: Cardiac imaging at high field (3T and above) suffers from image quality issues due to greater B1 + inhomogeneities and is frequently limited by SAR, which constrains the speed at which scans can be run without exceeding regulatory limits. Parallel transmission can mit...