Rupture of Extensively Porous Coated Stems. - A Case Series of 2 Patients

نویسندگان

  • X Gallart
  • G Claret
  • S Garcia
  • JA Fernández-Valencia
  • J Riba
چکیده

INTRODUCTION Mechanical failure of femoral stems of revision hip arthroplasty has been rarely reported. In the current study, the cause of two stem fractures, which occurred in vivo, was analysed with use of clinical and radiological data, and the functional result after revision is presented. CASE REPORT Two patients, A 70-year-old male and a 73-year-old female, both of Mediterranean ethnic, and both patients underwent a revision total hip replacement to an uncemmented extensively porous coated stem. Both stems suffered an implant fatigue in vivo at three years and at two years follow-up respectively. CONCLUSION Revision total hip arthroplasty is a procedure that will be performed more often the following years due to aging of population. Any orthopaedic surgeon performing hip surgery should be aware of the risk factors that can lead to total hip arthroplasty failure. In the analysed cases we can learn that the main factors related to this failure included the use of a small size stem (inferior to 14mm), an inadequate proximal osseous support because of trochanteric osteotomy, and a reduced preoperative bone stock. Although the use of cables has not been stated as a predisposing factor, we consider that they could also play a role in the development of this rare complication.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Revision Total Hip Arthroplasty Using an Extensively Porous Coated Femoral Stem

BACKGROUND To determine the benefit of an extensively porous coated femoral stem in patients receiving revision total hip arthroplasty. METHODS This study reviewed the results of 35 patients who received a revision total hip arthroplasty with extensively porous coated femoral stem between August, 1996, and December, 2002. The mean follow-up period was 77.5 months. The clinical and radiologica...

متن کامل

Bone density study of the proximal femur after hip arthroplasty with porous-coated implants

Progressive loss of bone mineral density around the femora! component of total hip replacement continues to pose a threat to long term prosthetic survival. A linear study was undertaken to measure bone mineral density on a monthly basis following total hip arthroplasty in 11 male patients. The opposite femur was used as the control measurement. Bone mineral density was unchanged at two mo...

متن کامل

Extensively porous-coated stems for femoral revision: reliable choice for stem revision in Paprosky femoral type III defects.

Between January 1999 and August 2008, ninety-six femoral revisions were performed with extensively porous-coated stems in Paprosky type III femoral defects (89 type IIIA and 7 type IIIB defects). Seven type IIIB defects with a mean canal of 16.5 mm were observed; 6 defects achieved stable bone ingrowth and 1 achieved stable fibrous condition. Average postoperative Harris Hip Score was 92.3±8 (r...

متن کامل

Minimally invasive hollow trephine technique is recommended for revision of broken uncemented and extensively porous-coated monolithic femoral stems: a review of three cases.

Revisions of fractured, uncemented and extensively porous-coated femoral stems are challenging. In this article, we report three cases of revision surgeries for broken, cementless, and extensively porous-coated femoral stems with Solution stems (DePuy, Warsaw, Indiana). In all cases, minimally invasive endofemoral revision was successfully performed without femoral osteotomy or cortical fenestr...

متن کامل

The use of uncemented extensively porous-coated femoral components in the management of Vancouver B2 and B3 periprosthetic femoral fractures.

We assessed the outcome of patients with Vancouver type B2 and B3 periprosthetic fractures treated with femoral revision using an uncemented extensively porous-coated implant. A retrospective clinical and radiographic assessment of 22 patients with a mean follow-up of 33.7 months was performed. The mean time from the index procedure to fracture was 10.8 years. There were 17 patients with a sati...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 4  شماره 

صفحات  -

تاریخ انتشار 2014