Pneumatic pulsatile ventricular assist devices in children under 1 year of age.
نویسندگان
چکیده
OBJECTIVE Although considerable progress has been made on ventricular assist devices (VAD) for adults, pneumatic pulsatile circulatory support in young infants is still limited. There is a need for long-term ventricular assist devices to bridge the failing myocardium of young children until recovery or transplantation. Miniaturized devices and innovative modalities need to be optimized. We report on our experience. METHODS From 1/1992 to 6/2004, 18 infants (6 male/12 female) under 1 year of age were treated with the Berlin Heart Excor VAD. The infants were divided into two groups, depending on the year of treatment. Group A consists of eight infants resuscitated and supported with a pulsatile pneumatic ventricular assist device between 1992 and 1998 and group B consists of 10 infants treated between 1999 and 6/2004. With the pediatric-sized Berlin Heart we used miniaturized extracorporeal pneumatically driven blood pumps, the lowest stroke volume being 10 ml. RESULTS In 18 children, age 3-345 (median 147) days, artificial replacement of heart function was applied for long-term support (1-64, median 10 days) as a life-saving measure in our hospital. Nine had LVAD and nine BVAD support. All were in cardiogenic shock with multiorgan failure; three had fulminant myocarditis, four cardiomyopathy, and one chronic stage of congenital heart disease. Five children were weaned from the system, three reached heart transplantation, and 10 died on the VAD. There were no differences between groups A and B regarding age, body weight or diagnosis, but the duration of mechanical support differed: Group A, median 2, range 1-16 days; group B, median 12, range 1-100 days. Since 1999 (group B), the survival rate of our small infants has increased to 70% whereas none of the infants in group A survived to be discharged. CONCLUSIONS The outcome of VAD support in small infants is no longer inferior to that of adult support, now optimized cannulas, modified anticoagulation and optimized surgical and intensive care management have been established.
منابع مشابه
The National Heart, Lung, and Blood Institute Pediatric Circulatory Support Program.
Options for the circulatory support of pediatric patients under the age of 5 years are currently limited to short-term extracorporeal devices, the use of which is often complicated by infection, bleeding, and thromboembolism. Recognizing this void, the National Heart, Lung, and Blood Institute solicited proposals for the development of novel circulatory support systems for infants and children ...
متن کاملLeft ventricular assist devices decrease fixed pulmonary hypertension in cardiac transplant candidates.
OBJECTIVE Fixed pulmonary hypertension is a contraindication for cardiac transplantation because of the increased risk of donor heart failure. We sought to determine whether left ventricular assist devices improve fixed pulmonary hypertension in cardiac transplant candidates to enable safe cardiac transplantation. METHODS Thirty-five consecutive cardiac transplant candidates (age 56 +/- 6 yea...
متن کاملMechanical circulatory support in patients of advanced age.
AIMS In recent years, non-pulsatile assist devices have been increasingly used for long-term circulatory support. However, it remains to be established whether pulsatile and non-pulsatile systems can ensure a similarly low rate of complications for extended periods of time in elderly patients. METHODS AND RESULTS Of 174 patients implanted between June 1991 and January 2009 [mean age: 65 +/- 3...
متن کاملClinical experience with the MEDOS HIA-VAD system in infants and children: a preliminary report.
BACKGROUND The need of pediatric cardiac assist is growing because of the complexity of the congenital conditions operated on and the increasing number of pediatric transplantations. We evaluated the newly developed pediatric MEDOS HIA-VAD ventricular assist device. METHODS The pneumatic paracorporeal ventricular assist device has three left ventricular sizes (10-, 25-, and 60-mL maximum stro...
متن کاملAdvanced heart failure treated with continuous-flow left ventricular assist device.
BACKGROUND Patients with advanced heart failure have improved survival rates and quality of life when treated with implanted pulsatile-flow left ventricular assist devices as compared with medical therapy. New continuous-flow devices are smaller and may be more durable than the pulsatile-flow devices. METHODS In this randomized trial, we enrolled patients with advanced heart failure who were ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery
دوره 28 2 شماره
صفحات -
تاریخ انتشار 2005