Validation of an integrated patient positioning system: Exactrac and iViewGT on Synergy Platform
نویسندگان
چکیده
Purpose: Evaluation of the newly integrated system for its validation and designing a quality assurance frame work to assess its geometrical, radiological and mechanical accuracy. Methods: Isocentric accuracy of two independent imaging modalities, kV based ExacTrac and MV based iViewGT was evaluated using Winston-Lutz test. A pelvic humanoid phantom was used for the radiological end-to-end test for its clinical utilization. Image quality for the systems was evaluated using Las Vegas Phantom and ETR-1 plate. The kV system was also assessed for kVp accuracy, kVp dose linearity, mAs-dose linearity and timer linearity and its accuracy. The system was tested for total filtration and output consistency. Tests for uniformity and noise measurement of kVp accuracy and its reproducibility, linearity test between applied kVp and the x-ray dose, linearity Test between applied mAs and the x-ray dose were also done. Results: Winston-Lutz test gave the isocentric deviation of 0.058 ± 0.015 mm with the average lateral deviation as 0.028 ± 0.021 mm, average longitudinal deviation as 0.032 ± 0.015 mm and average vertical deviation as 0.030 ± 0.016 mm. With the phantom test, the minimum measured displacement of Exactrac positioning was 0.2 ± 0.3 mm, 0.0 ± 0.2 mm and 0.1 ± 0.3 mm in longitudinal, lateral and vertical directions respectively. In image quality test, visible smallest visible hole size seen by both Exactrac and EPID imaging system was 5 mm and can resolve 1.5 lp/mm or better. The image uniformity was found to be 132.9 ± 3.06 pixels for MV images and 139 ± 4.41 pixels for kV images with the associated noise of ≤1% both for 120 kV-20 mAs and 4 MV beam energy of ExacTrac and iViewGT respectively. The uniformity and noise test, measured pixel intensity values for various points on MV and kV images separately were found to agree within ± 1% with respect to the central axis pixel value. The kVp accuracy and its reproducibility were tested for kV imager only. The deviation of kVp was found to be than ± 1% and its precision was seen to be even lesser than ± 0.1%. Linearity test between applied kVp and the x-ray dose and applied mAs and x-ray dose were tested only for the ExacTrac. Both the coefficient of linearity for kVp as well as mAs was found to be < 0.1. Conclusion: It is feasible to install ExacTrac imaging system with an Elekta linear accelerator. Both the imaging systems were found to be compatible in terms of image quality test and isocentric accuracy and can be used for the patient imaging in the same Linear accelerator.
منابع مشابه
Evaluation of Six-Dimensional Cranial Target Positioning Accuracy in Two Different Immobilization Methods Using Exactrac System
Introduction: The aim of this study was to determine the accuracy of two different immobilization methods in patient positioning in cranial radiotherapy. The six-dimensional (6D) target localization accuracy of using a dedicated stereotactic mask was compared with that of a conventional head mask by the ExacTrac system. Material and Methods: ...
متن کاملSU-E-J-46: Use of Brainlab-ExacTrac IGRT with Dot-Decimal Electron Compensator(BolusECT) for Accurate Patient Positioning.
PURPOSE To Validate Brainlab-ExacTrac IGRT with Dot-Decimal electron compensator(BolusECT) for accurate patient positioning. METHODS Dot-Decimal electron compensators(BolusECT) are used to achieve conformal dose distributions and greater sparing to normal structures for cases with varying patient contours and treated with single electron beam.Setup for such patients is difficult to verify due...
متن کاملQuality Assurance and Commissioning of an Infrared Marker-Based Patient Positioning System for Frameless Extracranial Stereotactic Radiotherapy
Rapid advancements in imaging technology have led to remarkable improvements in identification and localization of tumors, ushering the era of high-precision techniques in contemporary radiotherapy practice. However, uncertainties in patient set-up and organ motion during a course of fractionated radiotherapy can compromise precision of radiation therapy. Excellent accuracy has been achieved wi...
متن کاملQuantitative evaluation of patient setup uncertainty of stereotactic radiotherapy with the frameless 6D ExacTrac system using statistical modeling
The purpose of this study is to evaluate patient setup accuracy and quantify indi-vidual and cumulative positioning uncertainties associated with different hardware and software components of the stereotactic radiotherapy (SRS/SRT) with the frameless 6D ExacTrac system. A statistical model is used to evaluate positioning uncertainties of the different components of SRS/SRT treatment with the Br...
متن کاملThe spatial accuracy of two frameless, linear accelerator‐based systems for single‐isocenter, multitarget cranial radiosurgery
Single-isocenter, multitarget cranial stereotactic radiosurgery (SRS) is more efficient than using an isocenter for each target, but spatial positioning uncertainties can be magnified at locations away from the isocenter. This study reports on the spatial accuracy of two frameless, linac-based SRS systems for multitarget, single-isocenter SRS as a function of distance from the isocenter. One sy...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره شماره
صفحات -
تاریخ انتشار 2014