Design of a Selective and Sensitive PVC-Membrane Potentiometric Sensor for Strontium Ion Based on 1,10-Diaza-5,6-benzo-4,7-dioxacyclohexadecane-2,9-dione as a Neutral Ionophore
نویسندگان
چکیده
A novel PVC membrane sensor for the Sr ion based on 1,10-diaza-5,6-benzo-4,7dioxacyclohexadecane-2,9-dione has been prepared. The sensor possesses a Nernstian slope of 30.0 ± 0.6 mV decade over a wide linear concentration range of 1.6 × 10-3.0 ×10 M with a detection limit of 6.3 ×10 M. It has a fast response time of <15 s and can be used for at least two months without any considerable divergence in potential. The potentiometric response is independent of the pH of test solution in the pH range 4.3-9.4. The proposed electrode shows good selectivities over a variety of alkali, alkaline earth, and transition metal ions.
منابع مشابه
Novel ZrO2 ion-selective electrode based on 4,7,13,16,21,24-hexa oxa- 1,10- diazabicyc1o[8,8,8]-hexa cosane as neutral carrier in PVC matrix
In this work, a novel Zr02+ PVC-based membrane sensor based on 4,7,13,16,21,24-hexa oxa-1,10- diazabicyclo[8,8,8]- hexa cosane (HODBHC) as a new ionophore is presented. The sensor displays a linear dynamic rangebetween 1.0x 10-1 and 1.0x10 6 M, with a near Nernstian slope of 29.9 ± 0.4 mV per decade in an acidic solution(pH=1). The limit of detection is 8.0 x 1 0 7 M. The electrodes possess adv...
متن کاملHighly selective and sensitive sensor based on 4,13-didecyl-1,7,10,16-tetraoxa-4,13-diazacyclooctadecane as a heterocyclic ionophore
A new polyvinyl choloride (PVC) membrane sensor for cerium (III) was prepared based on 4,13-didecyl-1,7,10,16-tetraoxa-4,13-diazacyclooctadecane (kryptofix22DD) as an ionophore. The sensor exhibites a linear concentration range of 1.0×10-6to 1.0×10-1M, with a Nernstian slope of 18.0±1.0 mV/decade, and a detection limit of 9.7×10-7M. It has a fast response time (5 s) and can be used for two mont...
متن کاملChromium(III) ion selective electrode based on 2H-1,4-benzothioazine-2,3(4H)dione dioxime as a neutral carrier
A poly(vinyl chloride) membrane based on a new a-dioxime derivative as membrane carrier was prepared andinvestigated as a Cr(III) selective electrode. The electrode exhibits a good potentiometric response for Cr(III)over a wide concentration range 1.0 x 10-6 to 1.0 x 10-1 M with a slope 19.5+0.5 mV/decade and low detectionlimit of 8.9 x 10-7 M. It has a fast response time < 15 s. The best perfo...
متن کاملNano-level Monitoring of Yttrium by a Novel PVC-membrane Sensor Based on 2,9-dihydroxy-1,10-diphenoxy-4,7-dithiadecane
A poly (vinyl chloride)-based membrane of 2,9-dihydroxy-1,10-diphenoxy-4,7-dithiadecane (C20H26O4S2) as a neutral carrier was prepared and investigated as an Y-selective electrode. Effects of various plasticizers and anion excluders were studied in detail and improved performance was observed. The best performance was obtained for the membrane sensor having a composition of L: PVC: NPOE: PA in ...
متن کاملDetermination of Strontium Ion in Depilatory Powder, Toothpaste, and Raticide Samples using a Novel and Highly Sensitive Potentiometric Sensor Coupled with a Central Composite Design
The strontium ion, in water-soluble forms, is one of the major pollutant in drinking water. Therefore, a simple, rapid, selective, and sensitive potentiometric carbon paste electrode was developed to measure strontium ion in real samples with complex matrices. To prepare the potentiometric sensor, a synthetic ligand ((E)-4-(((2-amino-4-chlorophenyl)imino)methyl)-5-(hydroxymethyl)-2-methyl pyrid...