Irradiance of phototherapy equipment in maternity wards in Maceió.
نویسندگان
چکیده
The effectiveness of neonatal hyperbilirubinaemia treatment depends directly on the amount of energy emitted by light (irradiance). This cross-sectional study aimed to determine the irradiance of phototherapy equipment in maternity wards in Maceió, AL, Brazil. All equipment in use in the neonatal units in Maceió was included in the study, totaling 36 devices, except those in maintenance. The measurement of irradiance was carried out with a radiometer. We observed that 72.20% of the equipment presented efficient irradiance and 27.76% were inefficient. The conclusion is that the majority of phototherapy devices are emitting the minimum required irradiance for neonatal jaundice treatment.
منابع مشابه
[Standardizing the calibration of phototherapy devices - a proposal]
OBJECTIVE: To propose a method to evaluate and maintain the efficiency of equipment commonly used in Brazil in order to obtain uniform results in phototherapy treatments in different services. MATERIAL AND METHODS: The radiometer/photometer used to measure spectral irradiance is locally manufactured and fulfills basic requirements. Measure standardization seeks to be applicable and reproducible...
متن کاملEffect of sling application on efficacy of phototherapy in healthy term neonates with nonhemolytic jaundice: a randomized conrolled trial.
OBJECTIVE To evaluate the efficacy of white reflecting material (slings) hung from the sides of compact fluorescent lamp (CFL) phototherapy equipment in reducing the duration of phototherapy in healthy term neonates with non-hemolytic jaundice. DESIGN Randomized controlled trial. SETTING Postnatal ward of a tertiary level neonatal unit. PARTICIPANTS AND INTERVENTION Healthy term neonates ...
متن کاملThe importance of irradiance and area in neonatal phototherapy.
BACKGROUND Despite a long period of development, there are still considerable variations in the spectral output, the levels of irradiance, and irradiated area provided by commercial phototherapy systems. These variations depend on the types and output of the lamps used to produce the phototherapy, along with the design of the systems, and principally on whether the phototherapy is provided from...
متن کاملPrevention of neurodevelopmental sequelae of jaundice in the newborn.
Although its cause, jaundice in the newborn, is extremely common, the disabling neurological disorder kernicterus is very rare. Kernicterus may be prevented by selecting those infants who are at risk of extreme jaundice or who may be particularly vulnerable to bilirubin neurotoxicity. Because the tools for achieving that goal are inadequate, a secondary strategy is needed. This involves a plan ...
متن کاملPhototherapy and the use of heat shields in very low birthweight infants.
The irradiance produced by phototherapy units in use on a neonatal unit was measured. The phototherapy tubes were found to decline at a variable rate. Heat shields used to decrease heat and evaporative losses in the very low birthweight infant reduce irradiance and may appreciably limit the effectiveness of phototherapy.