چکیده
Percutaneous screw insertion for minimally displaced or reducible acetabular fracture using x-ray fluoroscopy and computer-assisted navigation system has been advocated by some authors. The purpose of this study was to compare intraoperative conditions and clinical results between isocentric C-arm 3-dimensional (Iso-C 3D) fluoroscopy and conventional fluoroscopy for percutaneous retrograde screwing of acetabular anterior column fracture.A prospective cohort study was conducted. A total of 22 patients were assigned to 2 different groups: 10 patients in the Iso-C 3D navigation group and 12 patients in the conventional group. The operative time, fluoroscopic time, time of screw insertion, blood loss, and accuracy were analyzed between the 2 groups.There were significant differences in operative time, screw insertion time, fluoroscopy time, and mean blood loss between the 2 groups. Totally 2 of 12 (16.7%) screws were misplaced in the conventional fluoroscopy group, and all 10 screws were in safe zones in the navigation group. Percutaneous screw fixation using the Iso-C 3D computer-assisted navigation system significantly reduced the intraoperative fluoroscopy time and blood loss in percutaneous screwing for acetabular anterior column fracture.The Iso-C 3D computer-assisted navigation system provided a reliable and effective method for percutaneous screw insertion in acetabular anterior column fractures compared to conventional fluoroscopy.
منابع مشابه
Accuracy and Safety in Pedicle Screw Placement in the Thoracic and Lumbar Spines : Comparison Study between Conventional C-Arm Fluoroscopy and Navigation Coupled with O-Arm® Guided Methods
OBJECTIVE The authors performed a retrospective study to assess the accuracy and clinical benefits of a navigation coupled with O-arm® system guided method in the thoracic and lumbar spines by comparing with a C-arm fluoroscopy-guided method. METHODS Under the navigation guidance, 106 pedicle screws inserted from T7 to S1 in 24 patients, and using the fluoroscopy guidance, 204 pedicle screws ...
متن کاملIntraoperative three-dimensional fluoroscopy-based computerized tomography guidance for percutaneous kyphoplasty.
OBJECT Percutaneous kyphoplasty is an established method for the treatment of pathological vertebral compression fractures (VCFs). This procedure is usually performed with the aid of biplanar fluoroscopic image guidance. There are currently no published clinical studies in which the use of three-dimensional (3D) image guidance to facilitate this technique has been evaluated. The purpose of this...
متن کاملAn Effective and Feasible Method, “Hammering Technique,” for Percutaneous Fixation of Anterior Column Acetabular Fracture
Objective. The objective of this study was to evaluate the effectiveness and advantages of percutaneous fixation of anterior column acetabular fracture with "hammering technique." Materials and Methods. We retrospectively reviewed 16 patients with percutaneous fixation of anterior column acetabular fracture with "hammering technique." There were 11 males and 5 females with an average age of 38....
متن کاملC-arm fluoroscopic cone beam CT for guidance of minimally invasive spine interventions.
BACKGROUND Isocentric C-arm fluoroscopic cone beam CT (CBCT) is a new technique for near real time 3-D volume imaging guidance of percutaneous interventional procedures. In combination with digital flat panel detectors, CBCT has high spatial resolution with isotropic voxel size, allowing for high resolution image reconstruction in any plane, including 3D rotational reconstructions. CBCT combine...
متن کاملClinical Results of Percutaneous Fixation of Pelvic and Acetabular Fractures: A Minimally Invasive Internal Fixation Technique
Background: The pelvic ring fractures (PRF) and acetabular fractures (AF) are among the major orthopedic injuriesassociated with high rates of morbidity and mortality. Open surgical stabilization is the standard treatment for themajority of these complications. Percutaneous minimally invasive surgical stabilization of the fractures has become anaccepted treatment method for th...