Low-dose thallium-201 protocol with a cadmium-zinc-telluride cardiac camera.
نویسندگان
چکیده
OBJECTIVES Thallium-201 is efficient for myocardial perfusion imaging, but leads to relatively high radiation exposure in patients. The purpose of this study was to compare the efficiency of low-dose thallium-201 imaging with cadmium-zinc-telluride (CZT) cameras with regular-dose thallium-201 imaging with conventional cameras. METHODS We prospectively studied 137 consecutive patients referred for stress myocardial perfusion imaging who had previously had a myocardial single-photon emission computed tomography with thallium-201. We injected at stress a low dose of thallium-201 (1.1 MBq/kg, 28 µCi/kg), performed a 5-7 min scan with a CZT camera (GE DNM 530c), and assessed redistribution imaging when the initial images were abnormal. We compared the CZT scan with the conventional dual-head tomographic camera scan taken previously with a regular dose of thallium-201. RESULTS The average delay between both scans was 22 months. The stress dose was 88 ± 16 (2.38 ± 0.43 mCi) versus 125 ± 13 MBq (3.38 ± 0.34 mCi; a 30% reduction). The time for camera acquisition was 6 versus 13 min (a 54% reduction). The myocardial counts were increased two-fold with CZT (mean: 446 Kcounts). The quality of CZT images was better in 69% of the cases. There were 59 artifacts with conventional cameras and 29 artifacts with CZT (P<0.01). The diagnostic agreement was calculated in patients without clinical or angiographic changes between both scans (115 patients) and was high (97%). The effective dose at stress was less than 12 mSv. High myocardial counting allowed for further decrease in the injected activity, leading to an effective dose as low as 8 mSv. CONCLUSION With reduced activities of thallium-201 and low effective doses, the CZT camera provides reliable, high-quality imaging.
منابع مشابه
Thallium-201 kinetics in nonischemic canine myocardium.
Myocardial thallium-201 (20TI) kinetics have not been precisely defined because reliable techniques to determine continuous activity in vivo have not been available. In this study, an implantable miniature cadmium telluride radiation detection device was inserted through the left ventricular apex, positioned against the endocardium, and used for continuous on-line monitoring of myocardial 201TI...
متن کاملNuclear myocardial perfusion imaging with a cadmium-zinc-telluride detector technique: optimized protocol for scan time reduction.
UNLABELLED We aimed at establishing the optimal scan time for nuclear myocardial perfusion imaging (MPI) on an ultrafast cardiac gamma-camera using a novel cadmium-zinc-telluride (CZT) solid-state detector technology. METHODS Twenty patients (17 male; BMI range, 21.7-35.5 kg/m(2)) underwent 1-d (99m)Tc-tetrofosmin adenosine stress and rest MPI protocols, each with a 15-min acquisition on a st...
متن کاملEvaluation of left ventricular diastolic function with a dedicated cadmium-zinc-telluride cardiac camera: comparison with Doppler echocardiography.
AIMS To evaluate the relationships between measures of left ventricular (LV) filling dynamics at cadmium-zinc-telluride (CZT) imaging and indexes of LV diastolic function at transthoracic echocardiography. METHODS AND RESULTS Two hundred and forty-seven patients underwent myocardial perfusion imaging at rest and after stress with a low-dose CZT protocol and a baseline transthoracic echocardio...
متن کاملSerial Myocardial Imaging after a Single Dose of Thallium-201
Although thallium-201 exercise scintigraphy has been established for the detection of myocardial ischemia and viability, little is known regarding the myocardial thallium-201 kinetics during angioplasty. Herein, we report a 77-year old man with angina pectoris, in whom serial myocardial imaging after a single dose of thallium-201 was helpful in identifying not only the culprit lesion and myocar...
متن کاملSemiconductor detectors allow low-dose-low-dose 1-day SPECT myocardial perfusion imaging.
UNLABELLED Cadmium zinc telluride (CZT) detectors with linear counting rate response enable count subtraction in sequential scanning. We evaluated whether count subtraction eliminated the need for higher activity doses in the second part of the 1-d stress-rest myocardial perfusion imaging (MPI) protocol. METHODS For 50 patients (mean age ± SD, 66 ± 12 y) with visually abnormal (n = 42) or equ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Nuclear medicine communications
دوره 33 5 شماره
صفحات -