Biomechanical Evaluation of a Novel Posterior Integrated Clamp That Attaches to an Existing Posterior Instrumentation for Use in Thoracolumbar Revision

نویسندگان

  • Patrick Senatus
  • Suresh Reddy Chinthakunta
  • Pedram Vazifeh
  • Saif Khalil
چکیده

STUDY DESIGN An in vitro biomechanical study. PURPOSE To evaluate the biomechanics of a novel posterior integrated clamp (IC) that extends on an already implanted construct in comparison to single long continuous bilateral pedicle screw (BPS) and rod stabilization system. OVERVIEW OF LITERATURE Revision surgery in the thoracolumbar spine often necessitates further instrumentation following a failed previous back surgery. Stability of these reconstructed constructs is not known. METHODS Six osteoligamentous T12-L5 calf spines were tested on a spine motion simulator in the following configurations: intact, four level constructs (T13-L4), three level constructs (L1-L4), and two level constructs (L2-L4), by varying the ratio between BPS and IC. A load control protocol of 8 Nm moments was applied at a rate of 1°/sec to establish the range of motion value for each construct in flexion-extension, lateral bending, and axial rotation. Statistical analysis was performed on raw data using repeated measures analysis of variance and significance was set at p<0.05. RESULTS On an average, the reduction in motion for the four level continuous pedicle screw and rod construct (67%) was similar to those extended with integrated clamps (64%). Furthermore, for three level and two level constructs, no significant difference was observed between continuous pedicle screw constructs and those revised with the integrated clamps (regardless of the ratio between BPS and IC). CONCLUSIONS The novel posterior IC showed equivalent biomechanical rigidity to continuous pedicle screw rod constructs in revision scenarios. Clinical studies on posterior rod adjunct systems are necessary to confirm these results.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

بررسی عوارض جراحی پولیواسکولیوز

  This is a retrospective study of 83 patients with polioscoliosis, treated surgically in Shafa hospital (1970-1997), and the results are as Follow: implant failure: 15 cases (18%), pseudoar throsis: 12 cases (14.5%) deep infection: 4 cases (4.5%) superficial infection: 3 cases (3.6%) cast sore: 17 cases (20.5%) and loss of correction: 28 cases (33.7%).   Implant failure was more common in Harr...

متن کامل

Radiological outcome of short segment posterior stabilisation and fusion in thoracolumbar spine acute fracture.

INTRODUCTION The optimal management of thoracolumbar spine fractures remains a matter of controversy. The current literature implies that the use of short-segment pedicle screw fixation may be inappropriate because of its high reported failure rate. The purpose of this study is to report the short-term results of thoracolumbar burst and compression fractures treated with short-segment pedicle i...

متن کامل

A biomechanical comparison of 360° stabilizations for corpectomy and total spondylectomy: a cadaveric study in the thoracolumbar spine

BACKGROUND To date, there has been no adequate biomechanical model that would allow a quantitative comparison in terms of stability/stiffness between a corpectomy with the posterior column preserved and a total spondylectomy with the posterior column sacrificed. The objective of this study was to perform a biomechanical comparison of 360° stabilizations for corpectomy and total spondylectomy, u...

متن کامل

Anterior versus modified combined instrumentation for burst fractures of the thoracolumbar spine: a biomechanical study in calves.

PURPOSE To compare stability after anterior instrumentation alone versus modified combined anterior and posterior instrumentation for burst fractures of the thoracolumbar spine in calves. METHODS Thoracolumbar spines of 10 calves were used. An axial compression force was applied on each specimen using a material-testing machine, until there was a burst fracture at T12 or L1. Five specimens we...

متن کامل

Radiographic Parameters in Diagnosis of Posterior Ligamentous Complex Injury

Background and Aim: Diagnosing the status of the posterior ligament complex (PLC) plays an essential role in the management of patients with thoracolumbar fractures. In this study, due to the inefficiency of existing imaging modalities in the accurate detection of PLC damage, we investigated the relevance of some imaging parameters to specific guidelines for rapid PLC injury detection. Methods...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 7  شماره 

صفحات  -

تاریخ انتشار 2013