Evaluation of two new immunochromatographic assays for diagnosis of malaria.
نویسندگان
چکیده
We assessed the performance of two new commercially available rapid diagnostic tests (RDTs) for malaria (SD Bioline Malaria Ag Pf test and Ag Pf/Pan test) in 200 patients with uncomplicated malaria between August and October 2007 in Madagascar. Results of the two RDTs were compared with those obtained by microscopy and real-time polymerase chain reaction. The sensitivity and specificity for detection of Plasmodium falciparum were 93% and 98.9%, respectively, for the SD Bioline Malaria Ag Pf test and 92.9% and 98.9% for the SD Bioline Malaria Ag Pf/Pan test. The sensitivity of the SD Bioline Malaria Ag Pf/Pan test was much lower for detection of other species (63.6%). The sensitivity of the two new assays decreased to 77.3% at parasitemia levels < 100 parasites/microL for detection of P. falciparum.
منابع مشابه
Assessment of a rapid immunochromatographic assay for the detection of avian influenza viruses
Rapid spreading of the low pathogenic avian influenza virus (AIV) caused by the H9N2 subtype and the highly pathogenic AIV caused by H5N1 have caused serious economic losses in the poultry industries of Asia. Therefore, the early detection of AIVs is crucial for the control of the disease. In the present study, the applicability of a rapid immunochromatographic (RIC) assay, which specifically d...
متن کاملPlaystation thumb--a new epidemic in children.
412 1. World Health Organization. New Perspectives, Malaria Diagnosis. Geneva: WHO, 2000. 2. Risch L, Bader M, Huber AR. Self-use of rapid tests for malaria diagnosis. Lancet 2000; 355: 237. 3. Mishra B, Samantaray JC, Kumar A, Mirdha BR. Study of false positivity of two rapid antigen detection tests for diagnosis of Plasmodium falciparum malaria. J Clin Microbiol 1999; 37: 1233. 4. Grobusch MP...
متن کاملComparison of Rapid Immunochromatographic Assays (ict Malaria P.f. /p.v. Test and Optimal Test) with Microscopy for Detection of Malaria Parasites
This study was conducted to compare the ability of developed immunochromatographic assays, ICT malaria P.f./P.v. test and OptiMAL test with standard microscopy for the diagnosis of malaria parasites. Blood samples were obtained from 150 patients clinically suspected of having malaria. A total of 59 samples (39.3%) were positive by blood films while 64 (42.7%) were positive by ICT p.f./p.v. and ...
متن کاملField evaluation of the ICT malaria P.f/P.v immunochromatographic test for detection of Plasmodium falciparum and Plasmodium vivax in patients with a presumptive clinical diagnosis of malaria in eastern Indonesia.
In areas such as eastern Indonesia where both Plasmodium falciparum and Plasmodium vivax occur, rapid antigen detection tests for malaria need to be able to detect both species. We evaluated the new combined P. falciparum-P. vivax immunochromatographic test (ICT Malaria P.f/P.v.) in Radamata Primary Health Centre, Sumba, Indonesia, from February to May 1998 with 560 symptomatic adults and child...
متن کاملEvaluation of the Makromed dipstick assay versus PCR for diagnosis of Plasmodium falciparum malaria in returned travelers.
Microscopy has been the traditional reference standard for malaria diagnosis. However, difficulty in maintaining the required technical skills, particularly in areas where malaria is not endemic, has prompted the development of rapid nonmicroscopic diagnostic assays based on the detection of malaria parasite antigen in whole blood. In this study, we evaluated the performance of one such device,...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- The American journal of tropical medicine and hygiene
دوره 79 5 شماره
صفحات -
تاریخ انتشار 2008