Evaluation of Functional Characteristics of 4 Oscillatory Positive Pressure Devices in a Simulated Cystic Fibrosis Model.
نویسندگان
چکیده
BACKGROUND Oscillatory positive expiratory pressure (OPEP) is an airway clearance therapy that delivers positive pressure and air-flow oscillations during exhalation. This study described functional characteristic differences of 4 OPEP devices during an active exhalation in a simulated model. We hypothesized peak pressure (Ppeak), positive expiratory pressure (PEP), oscillatory frequency (f), and pressure amplitude will differ, depending upon the device used, device resistance setting, and time (repeated consecutive active exhalations through the device). METHODS The ASL 5000 was scripted to simulate pulmonary mechanics of a pediatric cystic fibrosis patient with moderate to severe lung disease. Airway resistance was standardized at 17.1 cm H2O/L/s, pulmonary compliance at 42.1 mL/cm H2O, active exhalation at 22 breaths/min, and tidal volume at 409 mL. Resistance settings for the Acapella, RC-Cornet, Flutter, and Aerobika were adjusted to low, medium, and high. Values for f, Ppeak, PEP, and pressure amplitude were recorded for 1 min and graphically displayed. RESULTS Significant effects for time, device, and resistance (P < .01) were noted for Ppeak, PEP, and pressure amplitude at each resistance level, demonstrating that the devices functioned differently as more than one repetition of a series of consecutive active exhalations are performed. Significant interaction effects for device, resistance level, and time indicate inconsistent output for Ppeak (P < .01), PEP (P < .01), and pressure amplitude (P < .01). Oscillatory f values fell within the respective manufacturers' operational parameters. The Aerobika provided the most consistent pressure amplitude across resistance settings and produced the highest mean pressure amplitude at medium and high resistance settings. CONCLUSIONS Statistically significant and clinically relevant variations in Ppeak, PEP, and pressure amplitude occurred between devices and within a device, as the resistance setting changed. The combination of device, time, and resistance settings affects OPEP device output for pressure, amplitude, and oscillatory frequency. Functional variations may impact therapeutic effectiveness, warranting additional study to determine clinical impact.
منابع مشابه
Positive Expiratory Pressure (PEP) versus Conventional Chest Physiotherapy in Pediatric Patients with Acute Exacerbation of Cystic Fibrosis
Background Pulmonary involvement is the main cause of mortality in cystic fibrosis (CF). Airway clearance techniques are non-pharmacological complement options for CF patients. The aim of this study was to evaluate the short-term outcome of airway cleaning treatment in patients with cystic fibrosis in a children's hospital. Materials and Methods This clinical trial study conducted on 40 CF pati...
متن کاملInvestigation of Growth of Microorganisms within the Spacers of Children with Cystic Fibrosis
Background and aim: Cystic fibrosis is a monogenic disorder in several systems. Since spacers used by these patients may be used several times a day, many of them do not know how to wash and use them, so these devices are good places for bacteria to grow. Therefore, this study was performed to identify and determine the relationship between the growth of microorganisms inside spacers and the ef...
متن کاملCompared to maximal current management standards, oscillating positive expiratory pressure devices have not been shown to improve clinically relevant outcomes in COPD patients with acute exacerbation
A critical appraisal and clinical application of Ambrosino N, Callegari G, Galloni C, Brega S, Pinna G. Clinical evaluation of oscillating positive expiratory pressure for enhancing expectoration in diseases other than cystic fibrosis. Monaldi Arch Chest Dis. 1995;50(4):269-275.
متن کاملIn vitro evaluation of positive expiratory pressure devices attached to nebulizers.
BACKGROUND Patients with cystic fibrosis perform airway clearance techniques and receive nebulized medications on a regular basis. Some positive expiratory pressure (PEP) devices allow concomitant administration of aerosol. I hypothesized that this practice alters the aerosol characteristics and patient dose. I compared the aerosol characteristics and patient dose of nebulized albuterol from 2 ...
متن کاملEvaluation of Antibiotic Resistance Trend among Cystic Fibrosis Patients: A Single Center Study from 2014-2019
Background: Antibiotic resistance in Cystic Fibrosis (CF) patients has been a serious issue in their survival. The objective of this study was to investigate the antibiogram trend in serial admissions among CF patients. Materials and Methods: In this retrospective descriptive-analytical study, from 2014 to 2019, 95 known cases of CF pati...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Respiratory care
دوره 62 4 شماره
صفحات -
تاریخ انتشار 2017