A Study on Performance and Safety Tests of Electrosurgical Equipment

نویسندگان

  • A. Tavakoli Golpaygani
  • M.M. Movahedi
  • M. Reza
چکیده

INTRODUCTION Modern medicine employs a wide variety of instruments with different physiological effects and measurements. Periodic verifications are routinely used in legal metrology for industrial measuring instruments. The correct operation of electrosurgical generators is essential to ensure patient's safety and management of the risks associated with the use of high and low frequency electrical currents on human body. MATERIAL AND METHODS The metrological reliability of 20 electrosurgical equipment in six hospitals (3 private and 3 public) was evaluated in one of the provinces of Iran according to international and national standards. RESULTS The achieved results show that HF leakage current of ground-referenced generators are more than isolated generators and the power analysis of only eight units delivered acceptable output values and the precision in the output power measurements was low. CONCLUSION Results indicate a need for new and severe regulations on periodic performance verifications and medical equipment quality control program especially in high risk instruments. It is also necessary to provide training courses for operating staff in the field of meterology in medicine to be acquianted with critical parameters to get accuracy results with operation room equipment.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A Study on Performance and Safety Tests of Electrosurgical Equipment

Introduction: Modern medicine employs a wide variety of instruments with different physiological effects and measurements. Periodic verifications are routinely used in legal metrology for industrial measuring instruments. The correct operation of electrosurgical generators is essential to ensure patient’s safety and management of the risks associated with the use of high and low frequency ele...

متن کامل

Why Should We Have a Periodic Safety and Performance Program for Medical Devices

Nowadays, more than 10,000 different types of medical devices can be found in hospitals.These devices used in medical centers and hospitals for monitoring and treatment of patients require periodic safety and performance checking in order to have confidence in their functioning and operation. Physicians need better accurate medical measurements in order to better diagnose diseases, monitor pati...

متن کامل

A Study on Performance and Safety Tests of Defibrillator Equipment

Introduction: Nowadays, more than 10,000 different types of medical devices can be found in hospitals. This way, medical electrical equipment is being employed in a wide variety of fields in medical sciences with different physiological effects and measurements. Hospitals and medical centers must ensure that their critical medical devices are safe, accurate, reliable and operational at the requ...

متن کامل

analysis of Parameter related to electrocution accidents based on evaluation of electrical safety performance in small-scale industries

Introduction: Electrocutions have a low frequency and severe consequences. The purpose of this study was to analyze the parameters related to electrical accidents, and evaluation of electrical safety performance of small-scale industries. Method: This study was conducted in 2019 on 232 small-scale industries. Electrical accidents over a 7-year period were studied. The tools used included elect...

متن کامل

Performance Evaluation of Diagnostic X-Ray Equipment Regarding the Hospital Size in the Republic of Korea

  Introduction: The Republic of Korea has developed a national standard based on which diagnostic X-ray equipment must be tested every 3 years. Accordingly, the performance of X-ray equipment used in all hospitals is evaluated by national certification bodies in compliance with the safety management regulations for X-ray equipment. However, if the equipment i...