Evaluation of the Triage PPY on-site testing device for the detection of dextropropoxyphene in urine.

نویسندگان

  • Alphonse Poklis
  • Justin L Poklis
  • Lisa D Tarnai
  • Ronald C Backer
چکیده

A new point-of-care colloidal metal immunoassay urine drugs-of-abuse testing device, the BIOSITE TRIAGE Plus Propoxyphene (TPP), was evaluated for the rapid detection of dextropropoxyphene (PPY) and/or its primary metabolite, norpropoxyphene (NP), in urine at a total PPY/NP concentration of 300 ng/mL or greater. This assay has been added to the Triage device that tests for commonly abused drugs. Adding to drug-free urine PPY and NP established the linearity of the TPP assay at concentrations of 40%, 80%, 120%, and 160% of the cut-off concentration. No significant cross-reactivity was found at 1.0 g/L for 32 drugs commonly encountered in emergency department admissions. Significant cross-reactivity was observed only with diphenhydramine and tricyclic antidepressants. TPP results from 160 urine specimens screened for PPY and/or NP were compared to those obtained by testing with DRI enzyme immunoassay, Emit II plus immunoassay, Abuscreen Online immunoassay and gas chromatography-mass spectrometry (GC-MS). There was a 98.8% agreement of positive or negative results between TPP and both the DRI and OnLine assays. The two discordant TPP results were due to concentrations of NP below the TPP minimum cross-reactivity value of 400 ng/mL. These two specimens yielded GC-MS NP concentrations of 262 and 359 ng/mL. These NP concentrations were within +/- 20% of the cross-reactivity cut-off value for NP for TPP, DRI, and Online. There was only an 88% agreement of positive or negative results between TPP and the Emit assay. Twenty urine specimens yielding PPY positive results when tested by TPP were negative by Emit testing. The discordant TPP results were due to poor cross-reactivity of Emit to NP. A 98.8% agreement of positive PPY results was observed between TPP and GC-MS. Discordant urines were found to contain PPY concentrations below the cut-off value of the assay. TPP was found to be an accurate device for the detection of PPY and NP in urine.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Detection of napropamide by microwave resonator sensor using carbon nanotube – polypyrrole- chitosan layer

This paper presents the design and fabrication of proximity coupled feed disk resonator coated with Multi Walled Carbon Nanotubes (MWCNTs) and Polypyrrole-Chitosan (PPy-CHI) layers as a napropamide sensor. Computer Simulation Technology (CST) microwave studio was used to obtain the best design of disk resonator and feed line position in 5 GHz resonant frequency. Also, MWCNTs - PPy-CHI layers we...

متن کامل

ارزیابی روش‌های الکتروایمونواسی و 100Clinitek- برای تشخیص میکروآلبومینوری

Evaluation of Electroimmunoassay and Clinitek-100 for Detection of Microalbuminuria B.A. Jalali Khanabadi PhD , H. Mozaffari Khosravi PhD Received: 12/10/05 Sent for Revision: 04/03/06 Received Revised Manuscript: 20/01/07 Accepted: 25/01/07 Background and Objective: Detection of microalbuminuria is an important laboratory test for evaluation of diabetic complications. A reliable me...

متن کامل

Preparation and characterization magnetic polypyrrole composite microspheres decorated with copper(II) as a sensing platform for electrochemical detection of Carbamazepine

With a facile solvothermal technique, Ssynthesis and application of Fe3O4@PPy–CuIIcomposite microspheres in the carbon ionic liquid matrix have been reported as highly sensitive sensors for voltammetric determination of Carbamazepine (CBZ). The morphology, crystal phase and structure of synthesized The nanocomposite structure was confirmed by routine methods such as transmission electron micros...

متن کامل

Preparation and characterization magnetic polypyrrole composite microspheres decorated with copper(II) as a sensing platform for electrochemical detection of Carbamazepine

With a facile solvothermal technique, Ssynthesis and application of Fe3O4@PPy–CuIIcomposite microspheres in the carbon ionic liquid matrix have been reported as highly sensitive sensors for voltammetric determination of Carbamazepine (CBZ). The morphology, crystal phase and structure of synthesized The nanocomposite structure was confirmed by routine methods such as transmission electron micros...

متن کامل

Evaluation of 8-hour and 12-hour urine collection for detection of proteinuria in hypertensive women, Shariati hospital, 2004 - 2005

Introduction: Hypertensive disorders are common during pregnancy. Combination of hypertension and proteinuria markedly increase prenatal morbidity and mortality. The gold standard method for proteinuria assessment is 24-hour urine collection. This survey is designed to determine the correlation between protein content of 8-hour and 24-hour urine collections. Methods: This cross-sectional study...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of analytical toxicology

دوره 28 6  شماره 

صفحات  -

تاریخ انتشار 2004