The development and evaluation of individualized templates to assist transoral C2 articular mass or transpedicular screw placement in TARP-IV procedures: adult cadaver specimen study
نویسندگان
چکیده
OBJECTIVES The transoral atlantoaxial reduction plate system treats irreducible atlantoaxial dislocation from transoral atlantoaxial reduction plate-I to transoral atlantoaxial reduction plate-III. However, this system has demonstrated problems associated with screw loosening, atlantoaxial fixation and concealed or manifest neurovascular injuries. This study sought to design a set of individualized templates to improve the accuracy of anterior C2 screw placement in the transoral atlantoaxial reduction plate-IV procedure. METHODS A set of individualized templates was designed according to thin-slice computed tomography data obtained from 10 human cadavers. The templates contained cubic modules and drill guides to facilitate transoral atlantoaxial reduction plate positioning and anterior C2 screw placement. We performed 2 stages of cadaveric experiments with 2 cadavers in stage one and 8 in stage two. Finally, guided C2 screw placement was evaluated by reading postoperative computed tomography images and comparing the planned and inserted screw trajectories. RESULTS There were two cortical breaching screws in stage one and three in stage two, but only the cortical breaching screws in stage one were ranked critical. In stage two, the planned entry points and the transverse angles of the anterior C2 screws could be simulated, whereas the declination angles could not be simulated due to intraoperative blockage of the drill bit and screwdriver by the upper teeth. CONCLUSIONS It was feasible to use individualized templates to guide transoral C2 screw placement. Thus, these drill templates combined with transoral atlantoaxial reduction plate-IV, may improve the accuracy of transoral C2 screw placement and reduce related neurovascular complications.
منابع مشابه
Accuracy Assessment of Using Rapid Prototyping Drill Templates for Atlantoaxial Screw Placement: A Cadaver Study
Purpose. To preliminarily evaluate the feasibility and accuracy of using rapid prototyping drill templates (RPDTs) for C1 lateral mass screw (C1-LMS) and C2 pedicle screw (C2-PS) placement. Methods. 23 formalin-fixed craniocervical cadaver specimens were randomly divided into two groups. In the conventional method group, intraoperative fluoroscopy was used to assist the screw placement. In the ...
متن کامل“Direct vision” operation of posterior atlantoaxial transpedicular screw fixation for unstable atlantoaxial fractures
BACKGROUND The posterior screw fixation in atlas via posterior arch and lateral mass, also called C1 "pedicle" screw, combined with C2 pedicle screw fixiation has shown better biomechanical stability in unstable atlantoaxial fractures. However, its popularization has to fulfill the limitation imposed by anatomical characteristics. The aim of this study was to explore the manipulation, effect, a...
متن کاملPlacement of pedicle screws in the human cadaveric cervical spine: comparative accuracy of three techniques.
STUDY DESIGN This investigation was conducted in two parts. In the first part, a morphometric analysis of critical cervical pedicle dimensions were measured to create guidelines for cervical pedicle screw fixation based on posterior cervical topography. In the second part of the study, a human cadaver model was used to assess the accuracy and safety of transpedicular screw placement in the suba...
متن کاملAccuracy of fluoroscopy versus computer-assisted navigation for the placement of anterior cervical pedicle screws.
STUDY DESIGN Randomized laboratory cadaver study. OBJECTIVE The objective of this study was to determine the accuracy of anterior transpedicular screw placement in the cervical spine using conventional fluoroscopy versus computer-assisted navigation. SUMMARY OF BACKGROUND DATA Traditionally, global cervical instability has required anterior and posterior fixation due to the superior biomech...
متن کاملMorphometric Analysis and Classification of the Cross-Sectional Shape of the C2 Lamina
A thorough understanding of the morphology of the lamina of the second cervical vertebra (C2) is important for safe C2 translaminar screw placement. Although anatomical characteristics of the C2 lamina have been widely documented, individual differences in morphology have not been addressed. The aim of this study was to morphometrically analyze the cross-sectional shape of the C2 lamina and cla...
متن کامل