Transcutaneous measurement of carbon dioxide tension during extended monitoring: evaluation of accuracy and stability, and an algorithm for correcting calibration drift.

نویسندگان

  • David J Berlowitz
  • Jo Spong
  • Fergal J O'Donoghue
  • Rob J Pierce
  • Douglas J Brown
  • Donald A Campbell
  • Peter G Catcheside
  • Ian Gordon
  • Peter D Rochford
چکیده

BACKGROUND When polysomnography is indicated in a patient with a presumed sleep disorder, continuous monitoring of arterial carbon dioxide tension (P(aCO(2))) is desirable, especially if nocturnal hypoventilation is suspected. Transcutaneous CO(2) monitors (P(tcCO(2))) provide a noninvasive correlate of P(aCO(2)), but their accuracy and stability over extended monitoring have been considered inadequate for the diagnosis of hypoventilation. We examined the stability and accuracy of P(tcCO(2)) measurements and the performance of a previously described linear interpolation technique designed to correct for calibration drift. METHODS We compared the P(tcCO(2)) values from 2 TINA TCM-3 monitors to P(aCO(2)) values from arterial blood samples obtained at the beginning, every 15 min of the first hour, and then hourly over 8 hours of monitoring in 6 hemodynamically stable, male, intensive care patients (mean age 46 ± 17 y). RESULTS Time had a significant (P = .002) linear effect on the P(tcCO(2))-P(aCO(2)) difference, suggesting calibration drift over the monitoring period. We found no differences between monitor type or interaction between time and monitor type. For the 2 monitors the uncorrected bias was 3.6 mm Hg and the limits of agreement were -5.1 to 12.3 mm Hg. Our linear interpolation algorithm improved the bias and limits of agreement to 0.4 and -5.5 to 6.4 mm Hg, respectively. CONCLUSIONS Following stabilization and correction for both offset and drift, P(tcCO(2)) tracks P(aCO(2)) with minimal residual bias over 8 hours of monitoring. Should future research confirm these findings, then interpolated P(tcCO(2)) may have an increased role in detecting sleep hypoventilation and assessing the efficacy of treatment.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Transcutaneous estimation of arterial carbon dioxide in intensive care. Which electrode temperature?

Transcutaneous and arterial carbon dioxide were measured simultaneously in 57 children (age range 10 days to 14.3 years) undergoing intensive care. All were haemodynamically stable at the time of study. Mean calibration time with 5 and 10% carbon dioxide was 43 (range 38-58) minutes and mean arterialisation time was 10.5 (range 3-30) minutes. Duplicate hourly arterial samples over a four hour p...

متن کامل

Transcutaneous carbon dioxide for short-term monitoring of neonates.

We studied transcutaneous pCO2 monitoring in 70 neonates, most of them premature with respiratory distress syndrome. Measurements were at 44 degrees C. Calibration drift was large in some instances. Least squares linear regression analyses of transcutaneous pCO2 (y) vs arterial pCO2 (x) in kilopascals showed, for all observations (n = 516), for one observation randomly selected from each patien...

متن کامل

Combined pulse oximetry/cutaneous carbon dioxide tension monitoring during colonoscopies: pilot study with a smart ear clip.

BACKGROUND We compared the accuracy and practicability of a new combined ear sensor device measuring pulse oximetry and transcutaneous carbon dioxide tension. METHODS Validation studies were done by comparing the results of the combined sensor with arterial blood gas measurements. In an observational part, monitoring data were obtained from 25 patients undergoing colonoscopy, sedated with mid...

متن کامل

Transcutaneous and capillary pCO2 and pO2 measurements in healthy adults.

Transcutaneous carbon dioxide and oxygen tensions (tc-pCO2 and tc-pO2) were measured in seven healthy adult volunteers during hyperventilation in atmospheric air and during CO2 inhalation. Three skin sensors were applied to each subject: an O2 electrode, a CO2 electrode, and a combined O2-CO2 electrode, each heated to 44 degrees C. We observed close correlation between tc-pCO2 and capillary-pCO...

متن کامل

The accuracy of non-invasive carbon dioxide monitoring: a clinical evaluation of two transcutaneous systems.

We determined the accuracy of two transcutaneous carbon dioxide monitoring systems (SenTec Digital Monitor with V-Sign Sensor and TOSCA 500 with TOSCA Sensor 92) for the measurement of single values and trends in the arterial partial pressure of carbon dioxide in 122 adult patients during major surgery and in 50 adult patients in the intensive care unit. One or several paired measurements were ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Respiratory care

دوره 56 4  شماره 

صفحات  -

تاریخ انتشار 2011